ICU Medical, Inc. recalls
22 notices naming ICU Medical, Inc. as the recalling firm, since November 27, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
ICU Medical, Inc.: CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes…
Hazard (FDA): Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or…
FDAMedical devices
ICU Medical, Inc.: Plum administration sets and 7 more
Hazard (FDA): Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence…
FDAMedical devices
ICU Medical Inc.: Hotline Fluid Warming Set
Hazard (FDA): ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
FDAMedical devices
ICU Medical, Inc.: ICU Medical Plum Duo Infusion Pump and 1 more
Hazard (FDA): In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
FDAMedical devices
ICU Medical, Inc.: Plum Solo Precision IV Pump and 2 more
Hazard (FDA): Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or…
FDAMedical devices
ICU Medical, Inc.: Oncology Kit w/60"
Hazard (FDA): Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy…
FDAMedical devices
ICU Medical, Inc.: Tego Connector: REF: 011-D1000
Hazard (FDA): Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear…
FDAMedical devices
ICU Medical, Inc.: ICU Medical Plum Duo Infusion Pump and 2 more
Hazard (FDA): Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at…
FDAMedical devices
ICU Medical Inc.: Pulsator Arterial Blood Sampling Kit and 1 more
Hazard (FDA): Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage…
FDAMedical devices
ICU Medical, Inc.: Plum Duo Infusion System
Hazard (FDA): ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible…
FDAMedical devices
ICU Medical, Inc.: LifeShield Drug Library Management
Hazard (FDA): Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they…
FDAMedical devices
ICU Medical, Inc.: LifeShield Drug Library Management
Hazard (FDA): For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do…
FDAMedical devices
ICU Medical, Inc.: ChemoLock Bag Spike and 37 more
Hazard (FDA): Port weld of drug transfer device may separate or break during use and potentially result in a leak.
FDAMedical devices
ICU Medical, Inc.: IV Administration sets. 23"
Hazard (FDA): IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or…
FDAMedical devices
ICU Medical Inc.: Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented… and 5 more
Hazard (FDA): Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
FDAMedical devices
ICU Medical, Inc.: Plum Duo Infusion System
Hazard (FDA): ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
FDAMedical devices
ICU Medical, Inc.: SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve
Hazard (FDA): Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which…
FDAMedical devices
ICU Medical, Inc.: ICU Medical Plum Duo and 1 more
Hazard (FDA): Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager…
FDAMedical devices
ICU Medical, Inc.: LifeShield Drug Library Management
Hazard (FDA): Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery…
FDAMedical devices
Mislabeled
ICU Medical, Inc.: POTASSIUM CHLORIDE Inj. and 1 more
Hazard (FDA): Labeling: Label Error on Declared Strength. Cases labeled POTASSIUM CHLORIDE 20 mEq, may contain flexible containers with overwrap mislabeled as 10 mEq. The correct dosage strength of 20 mEq is printed on the labeling…
FDAMedicines
ICU Medical, Inc.: ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device…
Hazard (FDA): Due to finished goods incorrectly packaged and labeled.
FDAMedical devices
ICU Medical, Inc.: Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum…
Hazard (FDA): ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are…
FDAMedical devices