What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
8 more codes are in the record below.
IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014
- Lot or code
- REF/UDI-DI/Lot: B33359/10887709007823/14070544, 14126963; B9213/10840619045859/14130197; B9732/10887709015033/14149636; B9733/10887709015040/14176779; 011-C7014/10840619005068/14029638, 14065820, 14110075, 14211300, 14129823
- FDA recall number
- Z-0154-2026
- Quantity
- 23203
Distribution (FDA): Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- ICU Medical, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.