What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear; may result in fluid leaks, occluded fluid flow, inability to inject/withdraw blood product, delayed/interrupted therapy, exposure to biological contaminants, air infused into the body
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
332 more codes are in the record below.
Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
- Lot or code
- UDI-DI: 00840619026059, 00840619006785. REF/Lot(Expiration): 011-D1000/14015455(5/1/2029), 14044313(6/1/2029), 14228823(12/1/2029), 14274294(1/1/2030); 01C-D1000/14171414(10/1/2029), 14304670(2/1/2030); 055-D1000/13772668(9/1/2028), 13778926(9/1/2028), 13791786(10/1/2028), 13794986(10/1/2028), 13797963(10/1/2028), 13801102(10/1/2028), 13806764(10/1/2028), 13812991(11/1/2028), 13822705(11/1/2028), 13828556(11/1/2028), 13833879(11/1/2028), 13838937(12/1/2028), 13867515(1/1/2029), 13875052(1/1/2029), 13876502(1/1/2029), 13876503(1/1/2029), 13882254(1/1/2029), 13882256(1/1/2029), 13887995(1/1/2029), 13887996(1/1/2029), 13894349(2/1/2029), 13894350(2/1/2029), 13903971(2/1/2029), 13903972(2/1/2029), 14045054(6/1/2029), 14045055(6/1/2029), 14049379(6/1/2029), 14049404(6/1/2029), 14063784(6/1/2029), 14063785(6/1/2029), 14071890(7/1/2029), 14071891(7/1/2029), 14071892(7/1/2029), 14087320(7/1/2029), 14090798(7/1/2029), 14110500(8/1/2029), 14131560(8/1/2029), 14144999(9/1/2029), 14145000(9/1/2029), 14150873(9/1/2029), 14150874(9/1/2029), 14150875(9/1/2029), 14170224(10/1/2029), 14170226(10/1/2029), 14179702(10/1/2029), 14179704(10/1/2029), 14179706(10/1/2029), 14208522(11/1/2029), 14208523(11/1/2029), 14212866(11/1/2029), 14212870(11/1/2029), 14216154(11/1/2029), 14216155(11/1/2029), 14224623(11/1/2029), 14248560(12/1/2029), 14248561(12/1/2029), 14274291(1/1/2030), 1427429291/1/2030), 14278257(1/1/2030), 14278258(1/1/2030), 14278259(1/1/2030), 14278260(1/1/2030), 14300970(2/1/2030),…
- FDA recall number
- Z-1111-2026
- Quantity
- 14,684,607 items
The text above is the beginning of the record; 5,404 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide - US Nationwide distribution including in the states of FL, AZ, RI, TN, KS, NJ, NY, VA, SC, GA, CA, MS, TX, OH, NC, MA, WA, IL, PA, NV, NH, LA, MI, NE, ME, MN, NM, OK, MD, IA, OR, SD, WI, MO, AL, WY, IN, KY, DC, UT, AR, WV, AK, ND, CO and the countries of BE, DE, CO, ES, AU, AR, SA, SI, IT, LK, MY, AT, FI, GB, TW, MX, NL, IE, CA, GR, ZA, SE, SK, NZ, LT, CH, HR, FR, JO, PE, SG, BR, PL, RO, HU, CZ, NO, AE, EE, CN, MA, HK.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- ICU Medical, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.