What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Plum Solo Precision IV Pump, 40001-0401
- Lot or code
- UDI-DI M335400011 Plum Solo Precision IV Pump v1.0.3
- FDA recall number
- Z-2311-2026
- Quantity
- 13,613
Plum Duo Infusion Pump, 40002-0401
- Lot or code
- UDI-DI M335400021 Plum Duo Infusion Pump v1.2.3
- FDA recall number
- Z-2312-2026
- Quantity
- 13,613
Plum Duo Precision IV Pump, 40002-0403
- Lot or code
- UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3
- FDA recall number
- Z-2313-2026
- Quantity
- 13,613
Distribution (FDA): Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- ICU Medical, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.