What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
iLet Bionic Pancreas, REF: BB1001
- Lot or code
- iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)
- FDA recall number
- Z-2513-2026
- Quantity
- 4580
Distribution (FDA): US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beta Bionics, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.