What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
iLet Bionic Pancreas, REF: BB1001
- Lot or code
- Software versions 1.4.3, and 1.4.2
- FDA recall number
- Z-1996-2026
- Quantity
- 1,080
Distribution (FDA): US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beta Bionics, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.