Beta Bionics, Inc. recalls
3 notices naming Beta Bionics, Inc. as the recalling firm, since April 22, 2026. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Beta Bionics, Inc.: iLet Bionic Pancreas
Hazard (FDA): The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the…
FDAMedical devices
Beta Bionics, Inc.: iLet Bionic Pancreas
Hazard (FDA): insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal…
FDAMedical devices
Beta Bionics, Inc.: iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
Hazard (FDA): it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active…
FDAMedical devices