FDAMedical devicesClass IIOngoing

FDA recall 98383: Intuitive Surgical, Inc.: Brand Name: da Vinci S and 3 more

FDA recall 98383 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Due to increased complaints for broken/frayed grip cables for reusable instruments.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

4 more codes are in the record below.

  • Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A

    Lot or code
    Lot Code: 420179: Monopolar Curved Scissors (Hot Shears) GTIN: 00886874111505 FDA Medical Device Number: D088920 Material Number: 420179-10 Batch / System Numbers: M10101102 SH0935 M10101201 SH0935 M10101202 SH1279 M10101203 SH0059 M10101204 USG294 M10101206 USG294 M10101206 SH1378 M10101214 USG022 M10101214 SH0035 M10101214 SH0495 M10110104 SH1279 M10110111 SH1378 M10110112 USG294 M10110113 SH0075 M10110121 SG704 M10110121 SH1279 M10110124 SH0495 M10110126 USG721 M10110126 SH1391 M10110127 USG721 M10110127 SH1279 M10110127 SH2144 M10110202 USG294 M10110202 SH0495 M10110203 SH2144 M10110208 SG704 M10110210 SH1378 M10110216 SH1368 M10110216 USG294 M10110216 SH1391 M10110217 SH0495 M10110217 SH1279 M10110222 SH1279 M10110228 SH0547 M10110228 USG721 M10110228 USG022 M10110228 SH1378 M10110228 SH1279 M10110301 SH1279 M10110303 SH0495 M10110304 USG294 M10110304 SH0729 M10110304 SH0761 M10110307 USG294 M10110308 USG294 M10110308 SH0035 M10110308 SH0495 M10110308 SH1279 M10110311 USG294 M10110311 SH2144 M10110315 USG294 M10110318 USG294 M10110318 SH0935 M10110318 SH1279 M10110323 USG022 M10110324 SH0059 M10110324 USG022 M10110324 SH0495 M10110328 USG294 M10110328 SH0935 M10110328 SH1279 M10110329 USG294 M10110329 SH0935 M10110331 SH0935 M10110405 SH1378 M10110413 USG294 M10110413 SH1279 M10110414 SG704 M10110414 SH0495 M10110420 SH1279 M10110425 SH0035 M10110426 SH0059 M10110428 USG022 M10110429 SG2026 M10110429 SH0035 M10110504 USG294 M10110512 USG294 M10110512 USG022 M10110512…
    FDA recall number
    Z-1696-2026
    Quantity
    86904

    The text above is the beginning of the record; 51,689 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA

    Lot or code
    Lot Code: 420189, Double Fenestrated Grasper GTIN: 00886874111581 FDA Medical Device Listing Number: D088920 Material Number: 420189-07 Batch / System Numbers: M10120911 SH1182 M10130103 SH1182 M10130116 SH1182 M10130124 SH1182 M10130422 SH1182 M10130731 SH1182 M10130918 SH1182 M10110112 SG704 M10110202 SG704 M10110318 SG704 M10111028 SG704 M20111221 SG704 M10120220 SG704 M10120419 SG704 M10130110 SG704 M10130422 SG704 M10130726 SG704 M10130802 SG704 M10130802 SH0547 M10130918 SH0547 M10110808 SH0059 M10111027 SH0059 M10111104 SH0059 M10121123 SH0059 M10121213 SH0059 M10130103 SH0059 M10130815 SH1557 M10130625 SH1390 M10130731 SH1390 M10130802 SH1390 M10130603 USG721 M10120911 SH0825 M10120727 SH1378 M10121026 SH1378 M10121105 SH1378 M10130121 SH1378 M10130207 SH1378 M10130218 SH1378 M10130322 SH1378 M10130315 SH1378 M10130422 SH1378 M10130603 SH1378 M10130625 SH1378 M10130726 SH1378 M10131010 SH1378 M10110309 SH0729 M10110415 SH0729 M10110621 SH0729 M10110812 SH0729 M10121009 SH0729 M10130103 SH0729 M10130124 SH1622 M10120125 SH1076 M10121113 SH1076 M10130426 SH1604 M10120719 SH1385 M10120727 SH1385 M10120822 SH1385 M10120829 SH2144 M10111104 SH0761 M10120327 SH1276 M10120522 SH1276 M10121026 SH1276 M10130514 SH1276 M10130426 USG518 M10130829 USG518 Material Number: 420189-09 Batch / System Numbers: M10131104 SH1182 M10140228 SH1182 M10140414 SH1182 N10140508 SH1182 N10140619 SH1182 N10140703 SH1182 N10140812 SH1182 N10140902 SH1182 N10141204 SH1182 N10150430 SH1182…
    FDA recall number
    Z-1697-2026
    Quantity
    2370

    The text above is the beginning of the record; 2,638 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Brand Name: da Vinci S, Si Mega Needle Driver Instrument Product Name: da Vinci S, Si Mega Needle Driver Instrument Model/Catalog Number: 420194 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA

    Lot or code
    Lot Code: 420194, Mega Needle Driver GTIN: 00886874111611 FDA Medical Device Listing Number: D088920 Material Number: 420194-08 Batch / System Numbers: M10120201 SH1182 M20120412 SH1182 M10130121 SH1570 M10130306 SH1570 M10110602 SH0547 M10110627 SH0547 M10121129 SH0059 M10121214 SH0059 M10130429 SH0059 M10130521 SH0059 M10121105 SH1557 M10130517 SH1837 M10121109 SH1390 M10130508 SH1390 M10131004 SH1390 M10130306 USG689 M10130412 USG689 M10110302 SH1368 M10110614 SH1368 M10110620 SH1368 M10110817 SH1368 M10120705 SH1368 M10121022 SH1368 M10130429 SH1368 M10130509 SH1368 M10130722 SH1368 M10110131 USG294 M10110511 USG294 M10110616 USG294 M10110811 USG294 M10111003 USG294 M20111103 USG294 M10111206 USG294 M20120103 USG294 M10120130 USG294 M10120305 USG294 M10120425 USG294 M10120713 USG294 M10121004 USG294 M10121017 USG294 M10121022 USG294 M10121105 USG294 M10121107 USG294 M10121206 USG294 M10121207 USG294 M10130108 USG294 M10130112 USG294 M10130201 USG294 M10130213 USG294 M10130215 USG294 M10130529 USG294 M11130608 USG294 M10130718 USG294 M10130829 USG294 M10120321 SG2026 M10120605 SG2026 M10111114 SH0825 M10110330 SH0935 M10110401 SH0935 M10120123 SH0935 M10120313 SH0935 M10120404 SH0935 M10120802 SH0935 M10120919 SH0935 M10121121 SH0935 M10110302 SH0729 M10110503 SH0729 M10110701 SH0729 M10110809 SH0729 M10110817 SH0729 M10110825 SH0729 M10110913 SH0729 M10110923 SH0729 M10111116 SH0729 M20111123 SH0729 M10111201 SH0729 M10120120 SH0729 M10120130 SH0729 M10120210 SH0729…
    FDA recall number
    Z-1698-2026
    Quantity
    7819

    The text above is the beginning of the record; 8,142 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA

    Lot or code
    Lot Code: 420278: Grasping Retractor GTIN: 00886874111772 FDA Medical Device Listing Number: D088920 Material Number: 420278-04 Batch / System Numbers: M10110613 SH1182 M10120127 SH1182 M10120217 SH1182 M10120829 SH1182 M10130103 SH1182 M10130114 SH1182 M11130404 SH1182 M10130527 SH1182 M10130606 SH1182 M10130909 SH1182 M10130925 SH1182 S10101229 SG704 S10110105 SG704 M10121214 SH0547 M10110118 SH0059 M10110613 SH0059 S10110819 SH0059 M10121122 SH0059 M10121214 SH0059 M10130114 SH0059 M10130220 SH0059 M10130319 SH0059 M11130404 SH0059 M10120912 SH1390 M10130730 USG721 M10130909 USG721 M10110307 SH0729 M10130214 SH0729 M11130404 SH0729 M10130617 SH0729 M10130628 SH0729 M10120127 SH1076 M10130307 SH1279 M10111028 SH0761 Material Number: 420278-06 Batch / System Numbers: M10140115 SH1182 N10140725 SH1182 N10140923 SH1182 N10150324 SH1182 N10150417 SH1182 N10150721 SH1182 N10150608 SH1182 N10150820 SH1182 N10151113 SH1182 N10151214 SH1182 N10151029 SH1182 N10160304 SH1182 N10151201 SH1182 N10141205 SH1182 N10160921 SH1182 N10161123 SH1182 N10170516 SH1182 N10170601 SH1182 N10170921 SH1182 N10171213 SH1182 N10171226 SH1182 N10180417 SH1182 N10140815 SG704 N10150210 SG704 N10150608 SG704 N10160203 SG704 N10170117 SG704 N10170921 SG704 N10170822 SG704 N10171002 SG704 N10171226 SG704 M10140306 SH0059 N10141205 SH0059 N10150605 SH0059 N10170213 SH0059 N10171213 SH0059 N10171117 SH1964 N10180417 SH1964 N10170310 USG294 M10140306 USG721 M10140319 USG721 N10170412 USG721 N10180117 USG721…
    FDA recall number
    Z-1699-2026
    Quantity
    2660

    The text above is the beginning of the record; 747 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

Source

Issued by
U.S. Food and Drug Administration
Firm
Intuitive Surgical, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Intuitive Surgical, Inc., Month: April 2026, Week 2026-W15. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to increased complaints for broken/frayed grip cables for reusable instruments.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98383 on April 8, 2026.