Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98569: Intuitive Surgical, Inc.: Brand Name: Da Vinci 5 Product Name: ASSY

FDA recall 98569 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Brand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A Product Description: Preface: This report under 21 CFR 806.10 is being submitted for the dV5 system tower, as the NIR Handheld Camera or SOCK is developed and integrated as part of the dV5 system. The affected component of the dV5 system tower is the NIR handheld camera, and is supplied by Schoelly as the manufacturer. Therefore, the product description in this report references the NIR handheld camera system. Product Description: The NIR handheld camera system is compatible for use with the da Vinci¿ 5 system to aid in surgery. The components of the NIR handheld camera system include a camera control unit and a light source, NIR handheld camera, and additional components that together, connect to the Tower of the da Vinci 5 system. When setting up the da Vinci 5 system, the image from the NIR handheld camera system is displayed on the Tower monitor. Component: Yes, the SOCK is the component and integrated into the dV5 system

    Lot or code
    Lot Code: Part Number: 380746 UDI: 00886874119662 FDA Medical Device Listing Number: D531324 (U.S.) Part Number: 470653-01 Serial Numbers: 1008007 1008008 1008009 1008010 1008011 1008012 1008013 1008015 1008016 1008017 1008020 1008021 1008022 1008024 1008100 1008101 1008102 1008104 1008105 1008107 1008108 1008109 1008112 1008113 1008114 1008115 1008116 1008117 1008997 1008998 1009902 1010029 1010033 1010034 1010036 1010272 1010296 1010297 1010318 1010319 1010449 1010450 1010451 1010496 1010497 1010498 1010499 1010501 1010581 1010618 1010619 1010620 1010621 1010633 1010634 1010635 1010636 1010638 1010640 1010642 1010643 1010644 1010645 1010646 1011333 1011334 1011335 1011336 1011337 1011338 1011340 1011362 1011363 1011364 1011366 1011370 1011371 1011372 1011374 1011377 1011378 1011379 1011380 1011381 1011382 1011383 1011384 1011386 1011388 1011389 1011390 1011391 1011392 1011393 1011394 1011395 1011396 1012056 1012057 1012058 1012059 1012060 1012062 1012063 1012064 1012065 1012066 1012068 1012069 1012070 1012071 1012073 1012075 1012076 1012078 1012079 1012080 1012081 1012082 1012084 1012085 1012284 1012998 1012999 1013001 1013002 1013003 1013004 1013005 1013007 1013008 1013009 1013010 1013011 1013012 1013013 1013015 1013017 1013019 1013108 1013109 1013110 1013111 1013112 1013113 1013114 1013115 1013116 1013117 1013118 1013119 1013120 1013121 1013122 1013124 1013125 1013126 1013127 1013129 1013130 1013132 1013491 1013492 1013493 1013494 1013496 1013497 1014027 1014028 1014029…
    FDA recall number
    Z-2310-2026
    Quantity
    1357 systems

    The text above is the beginning of the record; 9,622 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NH, NJ, NM, NV, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Belgium, France, Germany, Italy, Japan, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, South Korea, Spain, Switzerland, and United Kingdom

Source

Issued by
U.S. Food and Drug Administration
Firm
Intuitive Surgical, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Intuitive Surgical, Inc., Month: June 2026, Week 2026-W24. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98569 on June 10, 2026.

Other notices of Intuitive Surgical, Inc.