What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
41 more codes are in the record below.
da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45
- Lot or code
- UDI: 00886874119747, Serial: SQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420, SQ0422, SQ0425, SQ0427, SQ0428, SQ0432, SQ0433, SQ0436, SQ0443, SQ0447, SQ0448, SQ0451, SQ0454, SQ0457, SQ0458, SQ0462, SQ0464, SQ0469, SQ0470, SQ0662, SQ0727, SQ0740, SQ0761
- FDA recall number
- Z-1355-2026
- Quantity
- 47
Distribution (FDA): US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Intuitive Surgical, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.