Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98133: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: SMV DSX New Line and 3 more

FDA recall 98133 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

  • SMV DSX New Line, System, Tomography, Computed, Emission

    Lot or code
    UDI/DI Not applicable: All serial numbers in distribution
    FDA recall number
    Z-1044-2026
    Quantity
    7 units
  • SMV DSI, System, Tomography, Computed, Emission

    Lot or code
    UDI/DI Not applicable: All serial numbers in distribution
    FDA recall number
    Z-1045-2026
    Quantity
    3 units
  • SMV BodyTrack, System, Tomography, Computed, Emission

    Lot or code
    UDI/DI Not applicable: All serial numbers in distribution
    FDA recall number
    Z-1046-2026
    Quantity
    12 units
  • MAXXUS, System, Tomography, Computed, Emission

    Lot or code
    UDI/DI Not applicable: All serial numbers in distribution
    FDA recall number
    Z-1047-2026
    Quantity
    38 units

Distribution (FDA): US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece

Source

Issued by
U.S. Food and Drug Administration
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, Month: January 2026, Week 2026-W04. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98133 on January 21, 2026.

Other notices of GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING