Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
SMV DSX New Line, System, Tomography, Computed, Emission
- Lot or code
- UDI/DI Not applicable: All serial numbers in distribution
- FDA recall number
- Z-1044-2026
- Quantity
- 7 units
SMV DSI, System, Tomography, Computed, Emission
- Lot or code
- UDI/DI Not applicable: All serial numbers in distribution
- FDA recall number
- Z-1045-2026
- Quantity
- 3 units
SMV BodyTrack, System, Tomography, Computed, Emission
- Lot or code
- UDI/DI Not applicable: All serial numbers in distribution
- FDA recall number
- Z-1046-2026
- Quantity
- 12 units
MAXXUS, System, Tomography, Computed, Emission
- Lot or code
- UDI/DI Not applicable: All serial numbers in distribution
- FDA recall number
- Z-1047-2026
- Quantity
- 38 units
Distribution (FDA): US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING