Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97193: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Infinia and 1 more

FDA recall 97193 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

  • Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4 Nuclear Medicine system

    Lot or code
    Serial Numbers: Infinia I 16143 16149 16179 16315 16335 16353 16826 17339 17799 18420 18442 18604 18605 18611 18624 18627 18628 18629 18637 18639 18641 18644 18645 18647 18648 18649 18650 18651 18652 18653 18655 18656 18657 193S08002; Infinia I Hawkeye 1 15006 15007 15013 16001 16003 16004 16005 16006 16007 16008 16009 16011 16012 16013 16015 16016 16019 16021 16022 16024 16026 16027 16028 16030 16033 16037 16038 16039 16043 16044 16048 16049 16050 16051 16052 16054 16057 16059 16060 16063 16064 16065 16070 16073 16074 16079 16081 16082 16083 16084 16087 16091 16099 16107 16109 16114 16123 16126 16127 16132 16134 16137 16140 16141 16148 16151 16153 16154 16155 16158 16159 16163 16164 16166 16168 16169 16171 16172 16173 16175 16177 16178 16182 16186 16433; Infinia II 16045 16094 16115 16135 16170 16190 16191 16194 16196 16198 16206 16207 16208 16209 16211 16212 16213 16214 16218 16220 16221 16223 16224 16227 16228 16229 16234 16235 16239 16244 16246 16250 16259 16262 16263 16271 16279 16285 16296 16303 16304 16305 16310 16312 16313 16318 16319 16322 16323 16324 16326 16334 16336 16355 16356 16365 16370 16371 16372 16376 16382 16404 16408 16411 16412 16413 16415 16418 16419 16421 16425 16426 16428 16434 16437 16442 16446 16455 16456 16457 16460 16465 16466 16469 16470 16474 16475 16478 16483 16489 16491 16497 16499 16500 16504 16509 16513 16525 16526 16528 16531 16533 16538 16539 16541 16542 16553 16557 16559 16560 16564 16566 16576 16579 16583 16585 16587 16592 16608 16611…
    FDA recall number
    Z-2097-2025
    Quantity
    1822

    The text above is the beginning of the record; 9,621 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

    Lot or code
    Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502 ***Updated 9/9/26*** Additional Serial Nos. ELSCINT VARICAM V-1530 V-1696 Not Available/V2309 8125-17-16 343 2488 1958 Not Available/E5516APEX 8125-16-25 1617 1008 VG GANTRY ASSY 1144
    FDA recall number
    Z-2098-2025
    Quantity
    21

Distribution (FDA): Worldwide - US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, Month: July 2025, Week 2025-W29. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97193 on July 16, 2025.

Other notices of GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING