What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0290-2026
- Quantity
- 11 units
Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0291-2026
- Quantity
- 64 units
Elscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0292-2026
- Quantity
- 45 units
Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0293-2026
- Quantity
- 28 units
Elscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0294-2026
- Quantity
- 15 units
Elscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0295-2026
- Quantity
- 3 units
Maxicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0296-2026
- Quantity
- 37 units
Millenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0297-2026
- Quantity
- 1386 units
Millenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0298-2026
- Quantity
- 82 units
Millenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0299-2026
- Quantity
- 292 units
Millenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0300-2026
- Quantity
- 54 units
Optima NX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0301-2026
- Quantity
- 1 unit
Sopha Medical Vision (SMV) DSXi/DST-XLi,/ DST-XL/ DSTi; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0302-2026
- Quantity
- 205 units
Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
- Lot or code
- all serial numbers in distribution
- FDA recall number
- Z-0303-2026
- Quantity
- 102 units
Distribution (FDA): Worldwide
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.