What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
63 more codes are in the record below.
RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System
- Lot or code
- UDI: 0735000201031020201216. GTIN: 07350002010310. Serial Numbers: 10.1.0.613. Software Revision: RayStation 10B. Expiration Date: 2027-02-01.
- FDA recall number
- Z-1102-2026
- Quantity
- 0
RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System
- Lot or code
- Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916, 35000201063120220616. GTIN: 07350002010389, 07350002010433, 07350002010440, 07350002010631. Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116, 11.0.4.15. Software Revision: RayStation 11A, 11A SP1, 11A SP2, 11A SP3. Expiration Date: 2027-06-22.
- FDA recall number
- Z-1103-2026
- Quantity
- 10 units
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System
- Lot or code
- UDI: 0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201060020220620 and 0735000201057020221222. GTIN: 07350002010426, 07350002010495, 07350002010501, 07350002010600 and 07350002010570. Serial Numbers: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12, 12.3.0.119. Software Revisions: RayStation 11B, 11B SP1, 11B SP2, 11B SP3, 11B SPT1. Expiration Date: 2028-03-20.
- FDA recall number
- Z-1104-2026
- Quantity
- 22 units
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Version: RayStation 12A, RayStation 12A SP1, RayStation 12A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
- Lot or code
- Lot Code: UDI: 0735000201054920220616, 0735000201067920221007, 0735000201073020230913 GTIN: 07350002010549, 07350002010679, 07350002010730 Serial Numbers: 13.0.0.1547, 13.1.0.144, 13.1.1.89 Software Revisions: RayStation 12A, 12A SP 1, 12A SP2 Expiration Date: 2028-07-07
- FDA recall number
- Z-1105-2026
- Quantity
- 8
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Product Description: Radiation Therapy Treatment Planning System Component: No
- Lot or code
- Lot Code: UDI: 0735000201055620230630, 0735000201072320231213, 0735000201088420250129 GTIN: 07350002010556, 07350002010723, 07350002010884 Serial Numbers: 14.0.0.3338, 15.0.0.430, 15.1.3.10 Software Revisions: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Expiration Date: 2030-02-21
- FDA recall number
- Z-1106-2026
- Quantity
- 119
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStation 10B SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
- Lot or code
- Lot Code: UDI: 0735000201047120220131 GTIN: 07350002010471 Serial Numbers: 10.1.1.54 Software Revision: RayStation 10B SP1 Expiration Date: 2027-02-01
- FDA recall number
- Z-1107-2026
- Quantity
- 0
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0, 15.1.3 Software Version: RayStation 2024A SP1, RayStation 2024A SP2 Product Description: Radiation Therapy Treatment Planning System Component: No
- Lot or code
- UDI: 0735000201076120240508, 0735000201080820241217 GTIN: 07350002010761, 07350002010808 Serial Numbers: 15.1.0.852, 15.2.0.107 Software Revision: RayStation 2024A SP1, RayStation 2024A SP2 Expiration Date: 2030-04-15
- FDA recall number
- Z-1108-2026
- Quantity
- 0
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 Software Version: RayStation 2024B Product Description: Radiation Therapy Treatment Planning System Component: No
- Lot or code
- Lot Code: UDI: 0735000201077820240625 GTIN: 07350002010778 Serial Numbers: 16.0.0.847 Software Revision: RayStation 2024B Expiration Date: 2029-06-28
- FDA recall number
- Z-1109-2026
- Quantity
- 0
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
- Lot or code
- Lot Code: UDI: 0735000201082220250423, 0735000201094520250702 GTIN: 07350002010822, 07350002010945 Serial Numbers: 17.0.0.1270, 17.0.1.113 Software Revision: RayStation v2025, RayStation v2025 SP1 Expiration Date: 2030-10-22
- FDA recall number
- Z-1110-2026
- Quantity
- 0
Distribution (FDA): Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- RAYSEARCH LABORATORIES AB
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.