What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
121 more codes are in the record below.
Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1
- Lot or code
- Lot Code: UDI: 0735000201006820171130, 0735000201011220180608 and 0735000201013620180928 GTIN: 07350002010068, 07350002010112 and 07350002010136 Serial Numbers: 7.0.0.19, 8.0.0.61 and 8.0.1.10 Software Revision: RayStation 7, 8A and 8A Service Pack 1 Expiration Date: 2023-10-04
- FDA recall number
- Z-1185-2025
- Quantity
- 0
Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113 and 9.0.1.142 Software Version: RayStation 8B, 8B Service Pack 1 and 2, 9A and 9A Service Pack 1
- Lot or code
- Lot Code: UDI: 0735000201011220180608, 0735000201013620180928, 0735000201017420190612 and 0735000201048820220420 GTIN: 07350002010112, 07350002010136, 07350002010174 and 07350002010488 Serial Numbers: 8.0.0.61, 8.0.1.10, 9.0.0.113 and 9.0.1.142 Software Revisions: RayStation 8B, 8B Service Pack 1 and 2, 9A and 9A Service Pack 1 Expiration Date: 2027-12-15
- FDA recall number
- Z-1186-2025
- Quantity
- 1
Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52 and 10.0.2.10 Software Version: RayStation 9B, 9B Service Pack 1, 10A, 10A Service Pack 1 and 2
- Lot or code
- Lot Code: UDI: 0735000201026620191220, 0735000201029720200310, 0735000201030320200526, 0735000201036520200526 and 0735000201065520220608 GTIN: 07350002010266 and 07350002010297, 07350002010303, 07350002010365 and 07350002010655 Serial Numbers: 9.1.0.933, 9.2.0.483, 10.0.0.1154, 10.0.1.52 and 10.0.2.10 Software Revisions: RayStation 9B, 9B Service Pack 1, 10A, 10A Service Pack 1 and 2 Expiration Date: 2025-06-16
- FDA recall number
- Z-1187-2025
- Quantity
- 1
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1
- Lot or code
- Lot Code: UDI: 0735000201031020201216 and 0735000201047120220128 GTIN: 07350002010310 and 07350002010471 Serial Numbers: 10.1.0.613 and 10.1.1.54 Software Revisions: RayStation 10B and 10B Service Pack 1 Expiration Date: 2027-02-01
- FDA recall number
- Z-1188-2025
- Quantity
- 0
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3
- Lot or code
- Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916 and 0735000201063120220616 GTIN: 07350002010389, 07350002010433, 07350002010440 and 07350002010631 Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116 and 11.0.4.15 Software Revisions: RayStation 11A and 11A Service Pack 1, 2 and 3 Expiration Date: 2027-06-22
- FDA recall number
- Z-1189-2025
- Quantity
- 6
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Version: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1
- Lot or code
- Lot Code: UDI: 0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201060020220620 and 0735000201057020221222 GTIN: 07350002010426, 07350002010495, 07350002010501, 07350002010600 and 07350002010570 Serial Numbers: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Revisions: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1 Expiration Date: 2028-03-20
- FDA recall number
- Z-1190-2025
- Quantity
- 5
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Version: RayStation 12A, 12A Service Pack 1 and 2 and 2023B
- Lot or code
- Lot Code: UDI: 0735000201054920220616, 0735000201067920221007, 0735000201073020230913 and 0735000201055620230630 GTIN: 07350002010549, 07350002010679, 07350002010730 and 07350002010556 Serial Numbers: 13.0.0.1547, 13.1.0.144, 13.1.1.89 and 14.0.0.3338 Software Revisions: RayStation 12A, 12A Service Pack 1 and 2 and 2023B Expiration Date: 2028-07-07
- FDA recall number
- Z-1191-2025
- Quantity
- 32
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.1.1.2, 6.2.0.7, 6.3.0.6 Software Version: RayStation 6 Service Pack 1, 2 and 3
- Lot or code
- Lot Code: UDI: 0735000201008220170529, 0735000201007520170630 and 0735000201024220190923 GTIN: 07350002010082, 07350002010075 and 07350002010242 Serial Numbers: 6.1.1.2, 6.2.0.7 and 6.3.0.6 Software Revisions: RayStation 6 Service Pack 1, 2 and 3 Expiration Date: 2024-10-18
- FDA recall number
- Z-1192-2025
- Quantity
- 1
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.24 Software Version: RayStation 6
- Lot or code
- Lot Code: UDI: 00735000201001320161214 GTIN: 07350002010013 Serial Numbers: 6.0.0.24 Software Revisions: RayStation 6 Expiration Date: 2021-12-22
- FDA recall number
- Z-1193-2025
- Quantity
- 0
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 5.0.2.35 and 5.0.3.17 Software Version: RayStation 5 Service Pack 2 and 3
- Lot or code
- Lot Code: UDI: 0735000201002020170502 GTIN: 07350002010020 Serial Numbers: 5.0.2.35 and 5.0.3.17 Software Revisions: RayStation 5 Service Pack 2 and 3 Expiration Date: 2028-12-21
- FDA recall number
- Z-1194-2025
- Quantity
- 0
Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4 and 4.7.6.7 Software Version: RayStation 4.5.1, 4.7.2, 4.7.3, 4.7.4, 4.7.5 and 4.7.6
- Lot or code
- Lot Code: Serial Numbers: 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4 and 4.7.6.7 Software Revisions: RayStation 4.5.1, 4.7.2, 4.7.3, 4.7.4, 4.7.5 and 4.7.6 Expiration Date: 2025-03-17
- FDA recall number
- Z-1195-2025
- Quantity
- 0
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Belgium ENG, Canada, China, China APactron, Denmark, France, Germany, Hong Kong S.A.R. China, India, Italy, Japan, Korea, Netherlands, Norway, Poland, Singapore, Spain, SwitzerlandENG, Taiwan, Thailand, UK.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- RAYSEARCH LABORATORIES AB
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.