RAYSEARCH LABORATORIES AB recalls
4 notices naming RAYSEARCH LABORATORIES AB as the recalling firm, since February 26, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
RAYSEARCH LABORATORIES AB: RayCare 2024A SP1. Oncology Information Systems. and 1 more
Hazard (FDA): Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
FDAMedical devices
RAYSEARCH LABORATORIES AB: RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan… and 8 more
Hazard (FDA): Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material…
FDAMedical devices
RAYSEARCH LABORATORIES AB: RayStation with the following product descriptions: 1. RayStation 7 and 8 more
Hazard (FDA): DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
FDAMedical devices
RAYSEARCH LABORATORIES AB: Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System… and 10 more
Hazard (FDA): Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…