Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98050: Intuitive Surgical, Inc.: Brand Name: Da Vinci Product Name: da Vinci X

FDA recall 98050 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

206 more codes are in the record below.

  • Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System Model/Catalog Number: IS4000; IS4200; IS5000 Software Version: N/A Product Description: The da Vinci Xi, X, and da Vinci 5 surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument

    Lot or code
    Lot Code: UDI: 00886874114216 FDA Medical Device Listing Number: D531324 See attachment Affected Product List UDI: 00886874114216 FDA Medical Device Listing Number: D295931 See attachment Affected Product List UDI: 00886574119747 FDA Medical Device Listing Number: D531324 See attachment Affected Product List da Vinci Xi (U.S.) USM Serial Numbers / System Serial Numbers: 402229 SK0835 766250 SK2180 797416 SK3689 674366 SK0095 453952 SK0135 10200891 RSK8057 675618 SK3649 634899 SK2504 440133 SK0682 362922 SK6006 405421 SK0449 390728 SK0389 608999 SK2328 395840 SK0523 368190 SK0837 441220 SK0909 596758 SK5711 474567 SK0075 376348 RSK8009 468435 SK1448 586220 SK5354 417890 SK2763 620604 SK2121 454704 SK1314 601085 SK4517 360714 SK2497 502409 SK0480 450881 SK1012 374877 SK2478 614195 SK1155 10012820 SK2135 685557 SK2489 575597 SK5978 432486 SK3140 624939 SK4316 357017 SK0157 448370 SK0959 432785 SK1380 552202 SK4752 801777 SK3600 598141 SK1967 653902 SK0048 791279 SK5269 392895 SK0992 423997 SK1265 10130883 SK5682 10345574 SK0756 609316 SK0756 643879 SK1802 721374 SK2851 721385 SK2851 721390 SK2851 726207 SK2851 423127 SK0159 453587 SK0159 789074 SK0159 803938 SK3632 808928 SK3632 811477 SK3632 405752 SK0886 821001 SK3749 821012 SK3749 821018 SK3749 396587 SK0448 396835 SK0448 396956 SK0448 539536 SK0448 10201501 SK2573 364250 SK2573 389078 SK2573 10581505 SK8006 10591492 SK8006 10592026 SK8006 10592897 SK8006 10397285 SK7337 10398002 SK7337 10403504 SK7337 626074 SK7337 682234…
    FDA recall number
    Z-1232-2026
    Quantity
    219

    The text above is the beginning of the record; 2,175 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, Netherlands, South Korea, Spain, Switzerland, Taiwan and United Kingdom

Source

Issued by
U.S. Food and Drug Administration
Firm
Intuitive Surgical, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Intuitive Surgical, Inc., Month: February 2026, Week 2026-W07. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98050 on February 11, 2026.

Other notices of Intuitive Surgical, Inc.