FDA recall 97450: Maquet Cardiovascular, LLC: Heartstring III Proximal Seal System. Intravascular anastomosis occluder. and 2 more

FDA recall 97450 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

161 more codes are in the record below.

  • Heartstring III Proximal Seal System. Intravascular anastomosis occluder.

    Lot or code
    Model No. HS-3045; UDI: 00607567700307; Serial No. 3000365549, 3000428264, 3000447331, 3000460255, 3000365549, 3000390361, 3000420107, 3000428264, 3000447331, 3000460255, 3000475915.
    FDA recall number
    Z-2586-2025
    Quantity
    126 units (91 US, 35 OUS)
  • Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.

    Lot or code
    Model No. HSK-3038; UDI: 00607567700314¿; Serial No. 3000377764, 3000378278, 3000379222, 3000379987, 3000383569, 3000388598, 3000390377, 3000394076, 3000396875, 3000398149, 3000400694, 3000402932, 3000403566, 3000404026, 3000404996, 3000405699, 3000407693, 3000409172, 3000409619, 3000410091, 3000410583, 3000412000, 3000412963, 3000415599, 3000416141, 3000417121, 3000417711, 3000418572, 3000418772, 3000419866, 3000420311, 3000422320, 3000422757, 3000423694, 3000424888, 3000425725, 3000428927, 3000429846, 3000431259, 3000431934, 3000434091, 3000434862, 3000435356, 3000436032, 3000436446, 3000440202, 3000441040, 3000441162, 3000442134, 3000443538, 3000443910, 3000444136, 3000446374, 3000447343, 3000448483, 3000449214, 3000449897, 3000450561, 3000451914, 3000452532, 3000453432, 3000453804, 3000455728, 3000456346, 3000457969, 3000458434, 3000459087, 3000459889, 3000460034, 3000461587, 3000464463, 3000465229, 3000465676, 3000465679, 3000466111, 3000466878, 3000467559, 3000470706, 3000471492, 3000473348, 3000473350, 3000475032, 3000475033, 3000476158, 3000476611, 3000477574, 3000478410, 3000479294, 3000479465, 3000481661, 3000484284, 3000485833, 3000488966, 3000377764, 3000378278, 3000379222, 3000379987, 3000388598, 3000391049, 3000394076, 3000396875, 3000398149, 3000400694, 3000403566, 3000404026, 3000404996, 3000405699, 3000409172, 3000409619, 3000410091, 3000412000, 3000412963, 3000415599, 3000417121, 3000417752, 3000418572, 3000419866, 3000420311, 3000422320, 3000423694,…
    FDA recall number
    Z-2587-2025
    Quantity
    28,331 units (16,615 US, 11,716 OUS)

    The text above is the beginning of the record; 780 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Heartstring III Proximal Seal System, 4.3 MM. Intravascular anastomosis occluder.

    Lot or code
    Model No. HSK-3043; UDI: 00607567700321; Serial No. 3000379223, 3000382433, 3000385929, 3000391405, 3000399719, 3000404541, 3000406469, 3000408189, 3000411816, 3000414253, 3000414613, 3000415212, 3000419765, 3000421682, 3000422957, 3000427645, 3000428019, 3000429074, 3000438521, 3000442986, 3000444153, 3000444386, 3000451113, 3000451189, 3000454605, 3000457540, 3000459949, 3000459992, 3000465374, 3000466176, 3000466177, 3000471825, 3000473063, 3000473064, 3000474240, 3000474241, 3000475497, 3000478491, 3000478495, 3000482450, 3000485137, 3000488054, , 3000378620, 3000379223, 3000385929, 3000391405, 3000399719, 3000404541, 3000406469, 3000411816, 3000414253, 3000414613, 3000415212, 3000419765, 3000421682, 3000422957, 3000427645, 3000428019, 3000429074, 3000438076, 3000442986, 3000444153, 3000444386, 3000451113, 3000451189, 3000454605, 3000459949, 3000459992, 3000465374, 3000465479, 3000466177, 3000468296, 3000471825, 3000473064, 3000474240, 3000474241, 3000475497, 3000478491, 3000478495, 3000479466, 3000484006, 3000485137, 3000488054.
    FDA recall number
    Z-2588-2025
    Quantity
    12,471 units (6,565 US, 5,906 OUS)

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.

Source

Issued by
U.S. Food and Drug Administration
Firm
Maquet Cardiovascular, LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Maquet Cardiovascular, LLC, Month: September 2025, Week 2025-W39. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97450 on September 24, 2025.

Other notices of Maquet Cardiovascular, LLC