Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96439: Maquet Cardiovascular, LLC: The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in…

FDA recall 96439 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

48 more codes are in the record below.

  • The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm extended length endoscope. The Harvesting Cannula has four lumens which house the endoscope, C-Ring, distal lens washer tube, and Harvesting Tool for cutting and cauterizing vessel branches. The Harvesting Tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning the device, cutting, and cauterization are performed under endoscopic visualization. The device is intended for specific use with the VASOVIEW HEMOPRO Power Supply, VASOVIEW HEMOPRO 2 extension cable, and an adapter cable.

    Lot or code
    Model No. VH-4000 (Hemopro 2), VH-4001 (Hemopro 2 w/ Vasoshield); UDI-DI: 00607567700406 (Hemopro 2), 00607567700451 (Hemopro 2 w/ Vasoshield); Lot/Serial No. 3000436447, 3000437191, 3000439557, 3000439600, 3000439623, 3000439922, 3000440187, 3000440220, 3000440283, 3000440754, 3000440957, 3000440981, 3000440982, 3000441144, 3000441319, 3000441541, 3000441774, 3000442133, 3000442263, 3000442323, 3000442627, 3000442871, 3000442953, 3000442955, 3000443198, 3000443216, 3000443545, 3000443632, 3000443851, 3000444131, 3000444151, 3000444484, 3000445702, 3000445704, 3000445941, 3000446003, 3000446365, 3000446489, 3000446709, 3000446774, 3000447195, 3000447535, 3000447597, 3000447845, 3000448208, 3000448471, 3000448761, 3000449047, 3000449170, 3000449267, 3000450191, 3000450967.
    FDA recall number
    Z-1622-2025
    Quantity
    22,244 units

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Mexico, United Arab Emirates, Singapore, India, Australia, Canada, Brazil, Hong Kong, Japan, South Korea, New Zealand.

Source

Issued by
U.S. Food and Drug Administration
Firm
Maquet Cardiovascular, LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Maquet Cardiovascular, LLC, Month: April 2025, Week 2025-W18. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96439 on April 30, 2025.

Other notices of Maquet Cardiovascular, LLC