7 notices naming Maquet Cardiovascular, LLC as the recalling firm, since November 27, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Maquet Cardiovascular, LLC: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System
Hazard (FDA): Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its…
FDAMedical devices
Maquet Cardiovascular, LLC: Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW…
Hazard (FDA): Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft…
FDAMedical devices
Maquet Cardiovascular, LLC: Heartstring III Proximal Seal System. Intravascular anastomosis occluder. and 2 more
Hazard (FDA): Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate…
FDAMedical devices
Maquet Cardiovascular, LLC: The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in…
Hazard (FDA): There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while…
FDAMedical devices
Maquet Cardiovascular, LLC: Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the…
Hazard (FDA): Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
FDAMedical devices
Maquet Cardiovascular, LLC: VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000… and 1 more
Hazard (FDA): Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
FDAMedical devices
Sterility
Maquet Cardiovascular, LLC: Acrobat-i Positioner. Model Number C-XP-5000Z and 2 more
Hazard (FDA): Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
FDAMedical devices