Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96617: Belmont Instrument LLC: Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile and 2 more

FDA recall 96617 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

11 more codes are in the record below.

  • Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Hyperthermia Procedure kit (P/N 902-00045) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and HP Procedure Pack (P/N 902-00048P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).

    Lot or code
    Model Number: 902-00045. UDI-DI: 00896128002589. Lot Numbers: 20241005, 20241108
    FDA recall number
    Z-1674-2025
    Quantity
    158 sets
  • Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00037) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Outflow Patient Line (P/N 902-00039P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).

    Lot or code
    Model Number: 902-00037. UDI-DI: 00896128002213. Lot Numbers: 20241012, 20241107
    FDA recall number
    Z-1675-2025
    Quantity
    208 sets
  • Straight Inflow/Bifurcated Outflow Line Kit. Model Number: 902-00038. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line Kit (P/N 902-00038) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and Straight Inflow/Bifurcated Outflow Patient Line (P/N 902-00040P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).

    Lot or code
    Model Number: 902-00038. UDI-DI: 00896128002220. Lot Number: 20241013
    FDA recall number
    Z-1676-2025
    Quantity
    136 sets

Distribution (FDA): Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.

Source

Issued by
U.S. Food and Drug Administration
Firm
Belmont Instrument LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Belmont Instrument LLC, Month: May 2025, Week 2025-W19. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96617 on May 7, 2025.

Other notices of Belmont Instrument LLC