Belmont Instrument LLC recalls
4 notices naming Belmont Instrument LLC as the recalling firm, since February 12, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Belmont Instrument LLC: Allon 2001. Part Number: 200-00263
Hazard (FDA): An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the…
FDAMedical devices
Belmont Instrument LLC: 3.0 Liter Disposable Set
Hazard (FDA): Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
FDAMedical devices
Belmont Instrument LLC: Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile and 2 more
Hazard (FDA): Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
FDAMedical devices
Belmont Instrument LLC: Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2…
Hazard (FDA): Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set
FDAMedical devices