What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed to be used in general operation in hospital or alternate care environments to provide warmed blood and fluids to any patients e10 kg requiring warmed infusion. Part Number:903-00006
- Lot or code
- UDI-DI: 00896128002022 Lot Numbers: 20240910, 20241002, 20241006, 20241014
- FDA recall number
- Z-1099-2025
- Quantity
- 1,029 cases (12,348 sets)
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Australia, Canada, China, Japan, Malaysia, Poland, Switzerland, Taiwan (Province of China), Ukraine.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Belmont Instrument LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.