What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
- Lot or code
- Soft ware version 1.57.20, UDI: +B595BSM20500/$7V1.57.20/16D20231129Y
- FDA recall number
- Z-1069-2025
- Quantity
- 62 units
Distribution (FDA): Worldwide - US Nationwide distribution in the states of NY, TX, OH, MD and the country of India.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Murata Vios, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.