Why it was recalled
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
- Lot or code
- Soft ware version 1.49.02 UDI: +B595BSM20500/\$\$7V1.49.02/16D20200504S and Software version 1.52.00 UDI: +B595BSM20500/\$\$7V1.52.00/16D20210305K
- FDA recall number
- Z-0956-2025
- Quantity
- 275 units
Distribution (FDA): Worldwide - US Nationwide distribution in the states of NY, LA and the country of India.