Murata Vios, Inc. recalls
4 notices naming Murata Vios, Inc. as the recalling firm, since January 22, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Murata Vios, Inc.: Vios Monitoring System Bedside Monitor Model BSM2050
Hazard (FDA): Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.
FDAMedical devices
Murata Vios, Inc.: Vios Monitoring System Bedside Monitor
Hazard (FDA): A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. When a failure occurs, a Pulse Ox Sensor Off warning condition will be…
FDAMedical devices
Murata Vios, Inc.: muRata Vios Monitoring System Model 2050
Hazard (FDA): During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central…
FDAMedical devices
Murata Vios, Inc.: muRata Vios Monitoring System Model 2050
Hazard (FDA): During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the…
FDAMedical devices