Why it was recalled
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
Impella RP with SmartAssist; Product Number: 0046-0035;
- Lot or code
- Product Number: 0046-0035; UDI-DI: 00813502011869; Serial Numbers: 492884, 492882, 492883, 522056, 522054, 522055, 522106, 522107, 522110, 522111, 522105, 522109, 520014, 520017, 520018, 520019, 520020, 520021, 520022, 520023, 520747, 520748, 520743, 520744, 520746, 520739, 520740, 520664, 520742, 520660, 520661, 520663, 520666, 520668, 518497, 520659, 520665, 518500, 518501, 518496, 518494, 518499, 518502, 518503, 518963, 518960, 518961, 518962, 518959, 518958, 518955, 518956, 507392, 518954, 507386, 507389, 507390, 507391, 507385, 507384, 507393, 492683, 492682, 492681, 492680, 489923, 489921, 489919, 489918, 489916, 489917, 489915, 492677, 492679, 492674, 509155, 509156, 509157, 509151, 509154, 509149, 491454, 509148, 509150, 491456, 491457, 491458, 491455, 491460, 491461, 491462, 491463, 486724, 486717, 486718, 486721, 476036, 476038, 476034, 476035, 476033, 476032, 476031, 489894, 489892, 489893, 489889, 491433, 491437, 489886, 489887, 489888, 491431, 491432, 491436, 484301, 491435, 484300, 483987, 483986, 483984, 483985, 483979, 483981, 483988, 484302, 484294, 484295, 484296, 484298, 486727, 486725, 486726, 486729, 486728, 486732, 486733, 486734, 469495, 469494, 469492, 469493, 469497, 469496, 469488, 469489, 469490, 469491, 476257, 484772, 472427, 472429, 476254, 484773, 484774, 484775, 472430, 476253, 476255, 476250, 476256, 484769, 484771, 467632, 472081, 476259, 472073, 472074, 476251, 476258, 484767, 472075, 472077, 472076, 472078, 472079, 472432, 484776, 471719,…
- FDA recall number
- Z-0829-2025
- Quantity
- 1,560 units
The text above is the beginning of the record; 11,058 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Impella RP Flex with SmartAssist; Product Number: 1000323;
- Lot or code
- Product Number: 1000323; UDI-DI: 00813502012811; Serial Numbers: 551121, 551122, 551123, 551124, 551125, 551127, 551128, 551129, 537606, 537610, 537612, 537615, 537607, 537608, 537609, 537613, 537614, 548547, 548548, 548550, 548553, 548554, 548555, 548556, 548552, 548559, 548560, 548561, 548562, 548563, 548564, 548565, 549638, 549640, 549641, 549642, 549643, 548558, 546672, 546675, 546676, 546678, 546679, 546680, 547552, 547556, 547558, 547559, 547560, 546652, 546654, 546656, 546657, 546658, 546659, 547561, 549644, 549646, 549647, 546661, 547563, 547564, 547566, 547567, 547568, 547569, 547570, 547571, 546681, 547189, 547191, 547192, 547193, 547194, 547195, 548827, 548830, 548831, 548834, 548835, 548836, 543714, 543716, 543719, 543720, 543721, 543722, 547169, 547170, 547171, 547172, 547173, 547174, 547188, 548188, 548189, 548196, 545799, 545800, 545801, 545802, 545803, 545804, 548191, 548192, 548193, 548194, 545486, 545488, 545795, 545796, 545797, 529171, 529172, 529174, 529175, 529179, 545475, 545478, 545481, 545482, 545483, 545489, 545490, 545491, 545492, 545494, 539350, 539352, 539353, 539354, 539355, 543274, 543275, 543276, 543277, 543278, 543279, 543282, 529170, 543273, 543283, 543284, 543286, 544125, 544126, 544130, 539356, 539357, 539358, 539359, 539794, 539795, 539797, 539799, 544123, 544124, 544127, 538387, 538388, 538389, 539790, 539791, 539792, 539793, 538381, 538382, 538383, 538384, 538385, 538386, 543285, 543288, 543289, 543292, 535401, 535403, 535404, 535405,…
- FDA recall number
- Z-0830-2025
- Quantity
- 2,364 units
The text above is the beginning of the record; 17,687 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Domestic: Nationwide Distribution International: Canada
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Abiomed, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.