Skip to content
FDAMedical devicesClass IOngoing

FDA recall 99778: Abiomed, Inc.: Automated Impella Controller and 2 more

FDA recall 99778 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.

    Lot or code
    Product Code: 0042-0000-US, 0042-0000-UK, 0042-0000-EU, 0042-0000-IN, 0042-0000-CA; UDI:00813502010022, 00813502011296, 00813502011289, 00813502011418, 00813502011272; Serial No. IC1014, IC1029, IC1038, IC1043, IC1042, IC1061, IC1075, IC1096, IC1122, IC1190, IC1202, IC1203, IC1224, IC1231, IC1238, IC1269, IC1289, IC1446, IC1612, IC1741, IC1771, IC1852, IC1853, IC1708, IC1964, IC2071, IC2069, IC2154, IC2227, IC2314, IC2315, IC2029, IC2321, IC2428, IC2429, IC2521, IC2561, IC2562, IC2575, IC2664, IC2665, IC2691, IC2692, IC2739, IC2782, IC2840, IC2841, IC2848, IC2891, IC2976, IC2977, IC3009, IC3010, IC3154, IC3155, IC3244, IC3246, IC3342, IC3398, IC3347, IC3351, IC3512, IC3609, IC3714, IC3715, IC3772, IC3773, IC3806, IC3822, IC3823, IC3919, IC3975, IC3976, IC4041, IC4042, IC4045, IC4060, IC4250, IC4259, IC4295, IC4339, IC4616, IC4640, IC4641, IC4708, IC4831, IC4832, IC4884, IC5120, IC5151, IC5201, IC5056, IC5549, IC5605, IC5661, IC5662, IC5598, IC5720, IC5721, IC5745, IC5749, IC5750, IC5828, IC5984, IC6042, IC6321, IC6322, IC6328, IC6329, IC6359, IC6360, IC6403, IC6596, IC6597, IC6622, IC6623, IC6632, IC6637, IC6673, IC6825, IC6846, IC6834, IC6835, IC1387, IC1869, IC1909, IC2277, IC2278, IC2871, IC2872, IC3293, IC3294, IC3295, IC3674, IC3675, IC3784, IC3785, IC4741, IC5219, IC5220, IC5553, IC5911, IC5912, IC5926, IC5927, IC6289, IC6286, IC6287, IC6540, IC6583, IC6584, IC6585, IC6845, IC6925, IC7057, IC7074, IC7073, IC7441, IC7442, IC7591, IC1285, IC1416, IC3719, IC3720, IC3776,…
    FDA recall number
    Z-0208-2027
    Quantity
    428 units

    The text above is the beginning of the record; 2,104 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.

    Lot or code
    Product Code: 0042-0010-US, 0042-0010-UK, 0042-0010-EU, 0042-0010-IN; UDI: 00813502010985, 00813502010978, 00813502010961, 00813502011425; Serial No. IC6962, IC7004, IC7036, IC7170, IC7213, IC7214, IC7239, IC7241, IC7677, IC7787, IC7880, IC8105, IC8106, IC8295, IC8308, IC8315, IC8316, IC8335, IC8541, IC8543, IC11456, IC7937, IC9832, IC9833, IC5409, IC7798, IC7800, IC7946, IC8070, IC8337, IC8441, IC7825, IC8126, IC8630, IC8640, IC9804, IC11451, IC11497, IC8388, IC8751, IC9501, IC9989, IC10102, IC10103, IC10318, IC10589, IC10463, IC10979, IC11299, IC8164, IC9348, IC9500, IC9581, IC9648, IC9777, IC9793, IC9797, IC9915, IC9919, IC9922, IC10278, IC10302, IC10983, IC10984, IC10985, IC11174, IC11410, IC13231, IC13263, IC13264, IC13265, IC13681, IC13682, IC10040, IC10043, IC8401, IC8748, IC8980, IC8609, IC9178, IC9177, IC9324, IC9328, IC9557, IC9558, IC9744, IC9745, IC9746, IC9908, IC9921, IC9969, IC9970, IC10063, IC10064, IC10149, IC10150, IC10238, IC10239, IC10341, IC10366, IC10367, IC12407, IC12408, IC12409, IC13057, IC13060, IC13061, IC13064, IC8333, IC8494, IC12326, IC4952, IC5041, IC5158, IC5281, IC5283, IC5284, IC5403, IC5404, IC5413, IC5527, IC5588, IC5635, IC5636, IC5641, IC5643, IC5666, IC5667, IC5804, IC5805, IC5918, IC5921, IC6013, IC6343, IC6421, IC6422, IC6461, IC6463, IC6465, IC6777, IC6785, IC6944, IC6945, IC6946, IC7161, IC7233, IC7232, IC7235, IC7626, IC7738, IC7739, IC7792, IC7793, IC7826, IC7956, IC8012, IC8258, IC8423, IC8424, IC8499, IC8511, IC8723, IC8596,…
    FDA recall number
    Z-0209-2027
    Quantity
    179 units

    The text above is the beginning of the record; 133 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Impella Connect Optical AIC, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.

    Lot or code
    Product Code:0042-0040-US, 0042-0040-UK, 0042-0040-EU, 0042-0040-CA; UDI: 00813502011401, 00813502012484, 00813502012217, 00813502012958; Serial No. IC8024, IC7912, IC8179, IC8216, IC8349, IC8415, IC8418, IC8529, IC8617, IC8808, IC8852, IC8906, IC9081, IC9082, IC9198, IC9227, IC9657, IC9987, IC10097, IC10098, IC10294, IC10327, IC10530, IC10531, IC10383, IC10578, IC10607, IC10549, IC10557, IC10717, IC10722, IC10769, IC10879, IC10877, IC10900, IC10926, IC11012, IC11068, IC11158, IC11168, IC11147, IC11203, IC11353, IC13018, IC13602, IC13606, IC13609, IC13610, IC13605, IC13498, IC13509, IC13506, IC13512, IC13514, IC13515, IC13517, IC13519, IC13513, IC13516, IC13534, IC13536, IC13537, IC13538, IC13535, IC13456, IC13495, IC13485, IC13575, IC13578, IC13551, IC13554, IC13548, IC13552, IC13553, IC13556, IC13558, IC13543, IC13549, IC13555, IC13559, IC13562, IC13564, IC13557, IC13563, IC13649, IC13651, IC13589, IC13591, IC13592, IC13598, IC13599, IC13600, IC13597, IC13593, IC13607, IC13608, IC13571, IC13572, IC13573, IC13596, IC13622, IC13620, IC13621, IC10025, IC9990, IC10240, IC10241, IC10592, IC10601, IC11198, IC11640, IC11642, IC11645, IC11646, IC13696, IC11123, IC11282, IC11264, IC11266, IC10146, IC10180, IC10371, IC10396, IC10397, IC10721, IC10772, IC10828, IC10835, IC11119, IC11072, IC11878, IC11958, IC11967, IC11969, IC11978, IC11986, IC11992, IC11997, IC12260, IC12261, IC12446, IC12454, IC12458, IC12459, IC12544, IC12546, IC12566, IC12586, IC12595, IC12596, IC12666, IC12670,…
    FDA recall number
    Z-0210-2027
    Quantity
    293 units

    The text above is the beginning of the record; 1,253 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom

Source

Issued by
U.S. Food and Drug Administration
Firm
Abiomed, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Abiomed, Inc., Month: November 2026, Week 2026-W45. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99778 on November 4, 2026.

Other notices of Abiomed, Inc.