Abiomed, Inc.: Automated Impella Controller and 2 more
Hazard (FDA): Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge…
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system…
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller and 2 more
Hazard (FDA): The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
FDAMedical devices
Abiomed, Inc.: ABIOMED Impella RP Flex with SmartAssist Set and 12 more
Hazard (FDA): A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr…
FDAMedical devices
Abiomed, Inc.: Abiomed 14 Fr x 13 cm Low Profile Introducer Kit for Impella CP. Impella Set Product… and 3 more
Hazard (FDA): Potential for thrombus formation during prolonged use of the introducer.
FDAMedical devices
Abiomed, Inc.: Impella CP Set with SmartAssist. Product Codes: 1000080
Hazard (FDA): Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper…
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process…
FDAMedical devices
Abiomed, Inc.: Purge Cassette
Hazard (FDA): Increased risk of purge leaks with Generation 1 purge cassettes.
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
FDAMedical devices
Abiomed, Inc.: Impella RP with SmartAssist. Product Code: 0046-0035. and 2 more
Hazard (FDA): Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.
FDAMedical devices
Abiomed, Inc.: Impella 5.5 with SmartAssist¿ S2
Hazard (FDA): Device packaged in incorrect outer box carton.
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or…
FDAMedical devices
Abiomed, Inc.: Automated Impella Controller
Hazard (FDA): A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
FDAMedical devices
Abiomed, Inc.: Impella RP with SmartAssist and 1 more
Hazard (FDA): Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary…