Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 95408: Bard Peripheral Vascular Inc: The EVSRF catheter is a sterile

FDA recall 95408 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

172 more codes are in the record below.

  • The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulated to permanently eliminate the vein from blood circulation. The catheter has a long shaft for insertion into a vein, a handle with start/stop button, and a connector cable. The catheter shaft has a heating element that is energized by the digiRF radiofrequency generator, which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter. A button on the catheter (or a foot pedal attached to the generator) begins an automated treatment cycle of 20 seconds long at a set temperature of 120 ¿C, and the treatment stops automatically when complete. The catheter is used within a sterile operative field, while the generator remains outside the sterile operative field at all times.

    Lot or code
    Catalog Number: VC-10A2.5-6F-100 UDI does: (01)00858254006015(17)241009(10)65497423 (01)00858254006015(17)241006(10)66030181 (01)00858254006015(17)241011(10)66599182 (01)00858254006015(17)241017(10)66634308 (01)00858254006015(17)241020(10)66784768 (01)00858254006015(17)241026(10)66793709 (01)00858254006015(17)241028(10)66793710 (01)00858254006015(17)241103(10)66897727 (01)00858254006015(17)241113(10)66947733 (01)00858254006015(17)241117(10)67054619 (01)00858254006015(17)241130(10)67308611 (01)00858254006015(17)250322(10)69548248 (01)00858254006015(17)250419(10)69548249 (01)00858254006015(17)250423(10)69548250 (01)00858254006015(17)250427(10)69567357 (01)00858254006015(17)250511(10)69567358 (01)00858254006015(17)250515(10)70322501 (01)00858254006015(17)250607(10)70523716 (01)00858254006015(17)250523(10)70597994 (01)00858254006015(17)250525(10)70751594 (01)00858254006015(17)250606(10)70791660 (01)00858254006015(17)250613(10)71054985 (01)00858254006015(17)250620(10)71197454 (01)00858254006015(17)250703(10)71197455 (01)00858254006015(17)250801(10)71197574 (01)00858254006015(17)250622(10)71319092 (01)00858254006015(17)250630(10)71361455 (01)00858254006015(17)250712(10)71589682 (01)00858254006015(17)250721(10)71780480 (01)00858254006015(17)250724(10)71780481 (01)00858254006015(17)250821(10)71974718 (01)00858254006015(17)250803(10)72167973 (01)00858254006015(17)250821(10)72216631 (01)00858254006015(17)250809(10)72216758 (01)00858254006015(17)250810(10)72282293…
    FDA recall number
    Z-0371-2025
    Quantity
    222,373 catheters

    The text above is the beginning of the record; 7,446 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Australia, Singapore, Austria, Belgium, Czechia, Denmark, France, Germany, Greece, Ireland, Italy, Poland, Slovenia Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates, and South Africa.

Source

Issued by
U.S. Food and Drug Administration
Firm
Bard Peripheral Vascular Inc
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Bard Peripheral Vascular Inc, Month: November 2024, Week 2024-W47. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95408 on November 20, 2024.

Other notices of Bard Peripheral Vascular Inc