Bard Peripheral Vascular Inc recalls
7 notices naming Bard Peripheral Vascular Inc as the recalling firm, since November 20, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Bard Peripheral Vascular Inc: EnCor EnCompass Breast Biopsy and Tissue Removal System Console
Hazard (FDA): Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to…
FDAMedical devices
Bard Peripheral Vascular Inc: Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff…
Hazard (FDA): Due to damage to outer tray that can potentially compromise the sterile barrier
FDAMedical devices
Bard Peripheral Vascular Inc: Safe-T-Centesis 6 Fr Catheter Drainage Tray
Hazard (FDA): Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt…
FDAMedical devices
Bard Peripheral Vascular Inc: Venclose digiRF Generators
Hazard (FDA): Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use…
FDAMedical devices
Bard Peripheral Vascular Inc: Rotarex Atherectomy System and Instructions for Use
Hazard (FDA): Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could…
FDAMedical devices
Bard Peripheral Vascular Inc: Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM
Hazard (FDA): Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle…
FDAMedical devices
Bard Peripheral Vascular Inc: The EVSRF catheter is a sterile
Hazard (FDA): The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead…