What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
- Lot or code
- UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130
- FDA recall number
- Z-0887-2026
- Quantity
- 1240
Distribution (FDA): US Nationwide distribution in the states of CA, MN, IN, MI, LA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Bard Peripheral Vascular Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.