Why it was recalled
As currently written, the Owner’s Booklets/System Instructions for Use fails to emphasize that users must seek medical attention immediately if they receive an E-5 error code and are experiencing symptoms of high glucose.
The recalled product
FT. LAUDERDALE, FL) – February 6, 2026 – Trividia Health, Inc., announced today that it is initiating a labeling correction which requires a modification of the Owner’s Booklets/System Instructions for Use for all TRUE METRIX, TRUE METRIX AIR, TRUE METRIX GO, and TRUE METRIX PRO Blood Glucose Monit
Company announcement, as FDA publishes it
(FT. LAUDERDALE, FL) – February 6, 2026 – Trividia Health, Inc., announced today that it is initiating a labeling correction which requires a modification of the Owner’s Booklets/System Instructions for Use for all TRUE METRIX, TRUE METRIX AIR, TRUE METRIX GO, and TRUE METRIX PRO Blood Glucose Monitoring Systems (collectively, the “Products”) distributed in the United States, United Kingdom, Mexico, Australia, and the Caribbean.
Trividia is updating the E-5 Error Code in the “Messages” section of the Owner’s Booklets/System Instructions for Use to emphasize that users must seek medical attention immediately if they receive an E-5 error code and are experiencing symptoms of high glucose .
The system displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error. As currently written, the instructions could potentially lead to a delay in treatment if the user does not seek medical attention immediately when they receive an E-5 error code and are experiencing symptoms of high glucose. A delay in treatment may result in serious adverse health consequences or death, especially for users with very high blood glucose levels.
Since August 2014, when TRUE METRIX was launched globally, there have been 114 reported serious injuries, and one (1) death associated with the E-5 Error Code.
UPDATED E-5 INSTRUCTIONS for TRUE METRIX, TRUE METRIX AIR, and TRUE METRIX GO:
UPDATED E-5 INSTRUCTIONS for TRUE METRIX PRO:
If you have any questions relating to the Owner’s Booklets/System Instructions for Use update, please call Trividia Health Customer Care Department toll-free at 1-888-835-2723 Monday - Friday 8AM-8PM EST (excluding holidays) or e-mail trividia0126CC@trividiahealth.com or visit www.trividiahealth.com/E-5productnotice .
The correction affects all TRUE METRIX branded Blood Glucose Meters distributed in the United States, United Kingdom, Mexico, Australia and the Caribbean. This includes our cobranded products sold under store or distribution partner names. Please refer to the Product Notice located at www.trividiahealth.com/E-5productnotice for more information on the list of co-brand partners and affected product labeling.
The company is sending the Product Notice to impacted customers with instructions on what to do, to post the notice where products are stored/sold, and to forward the notice to device users, if possible. These customers include pharmacies, mail order companies, and distributors where the TRUE METRIX® meters are sold. Please refer to the Product Notice located at www.trividiahealth.com/E-5productnotice for more information.
You may continue to use the TRUE METRIX® Products . Products are not to be returned or replaced. This correction does not require removal of the Products from where they are used or sold. This labeling correction impacts the Owner’s Booklets/System Instructions for Use that accompany the meters at purchase, as well as the online labeling and help guides located on Trividia Health’s website. Trividia Health will notify users of additional mitigation strategies as needed.
Trividia Health has notified the U.S. Food and Drug Administration (FDA) of this action.
Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Complete and submit the report Online www.fda.gov/medwatch/report.htm
Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Patient safety is our top priority, and we apologize for any inconvenience this correction may cause you.
About Trividia Health
Trividia Health, Inc., is a global health and wellness company based in Fort Lauderdale, Florida and a leading developer, manufacturer and marketer of advanced performance products for people with diabetes. With products sold under TRUE and store brand labels, the company is the exclusive partner and supplier of affordable, high-quality blood glucose monitoring and health and wellness solutions for the world’s leading retail pharmacies, distributors and mail service providers. For more information, please visit: www.TrividiaHealth.com .
Link to Updated Press Release
TRUE METRIX Blood Glucose Monitoring Systems
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Trividia Health, Inc.