Why it was recalled
The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
75 more codes are in the record below.
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter Only/RE4095-40, HealthyAccents TRUE METRIX Kit/RE4097-01, Shopko TRUE METRIX Kit/RE4098-01, Henry Schein TRUE METRIX Kit/RE4099-40, Henry Schein TRUE METRIX Meter Only/RE4099-45, Millennium Pharmay Services TRUE METRIX Kit/RE4202-43, Kroger TRUE METRIX Kit/RE4203-01, TOPCARE TRUE METRIX Kit/RE4209-01, Foster and Thrive TRUE METRIX Kit/RE4211-01, TRUE METRIX Meter Only/RE4H01-00, TRUE METRIX Kit/RE4H01-01, TRUE METRIX Starter Kit/RE4H01-04, TRUE METRIX NFRS Meter Only/RE4H01-40, TRUE METRIX NFRS Starter Kit/RE4H01-43,…
- Lot or code
- Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE4011-00/87701426520, RE4011-01/87701426254, RE4019-01/713733069816, RE4023-01/21292007256, RE4025-01/93351018992, RE4027-01/21292007065, RE4031-01/41220001286, RE4051-43/20612479197231, RE4054-00/10939722447, RE4054-01/10939719447, RE4060-01/321130788669, RE4061-01/41520147745, RE4066-01/11822445610, RE4078-40/21292007614, RE4081-03/21292014797, RE4087-01/77890394113, RE4089-00/52569138557, RE4089-01/52569138526, RE4095-40/21292007041, RE4097-01/725439988933, RE4098-01/21292008901, RE4099-40/21292014902, RE4099-45/21292008697, RE4202-43/21292007454, RE4203-01/41260016028, RE4209-01/36800234529, RE4211-01/10939958693, RE4H01-00/21292006082, RE4H01-01/21292006051, RE4H01-04/21292008895, RE4H01-40/21292006068, RE4H01-43/21292006075, RE4H01-51/21292014940. International: RE4i01-00/N/A, RE4i01-01/N/A, RE4i01-02/N/A, RE4i03-02/N/A, RE4i03-11/N/A, RE4i05-00/N/A, RE4i07-00/N/A, RE4i23-00/21292010133, RE4i29-11/N/A, RE4i61-02/N/A, RE4i61-02THI/N/A, RE4i82-11/21292012205, RE4i82-12/21292009977. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX Owner's Booklet/ RE4TVH03 Rev 56 and prior, TRUE METRIX Owner's Booklet (English)/ RE4TVH35 Rev 51 and prior, TRUE METRIX Owner's Booklet (Spanish)/ RE4TVH35S Rev 51 and prior, McKesson Med Surg TRUE METRIX Self Monitoring Owners Booklet/ RE4SUN03 Rev 55 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/…
- FDA recall number
- Z-1545-2026
- Quantity
- 12,067,458
The text above is the beginning of the record; 1,143 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices. Description/Item: TRUE METRIX PRO Meter Only/RE4H01P-40, TRUE METRIX PRO Starter Kit/RE4H01P-43, PROCURE TRUE METRIX PRO Meter Only/RE4212P-00, Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40, Henry Schein TRUE METRIX PRO Kit/RE4099P-43, McKESSON TRUE METRIX PRO Meter Only/RE4051P-00, Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.
- Lot or code
- Device Item/UDI-DI: RE4H01P-40/21292006099, RE4H01P-43/21292006105, RE4212P-00/10840330705681, RE4099P-40/21292014919, RE4099P-43/21292008673, RE4051P-00/20612479197217, RE4068P-01/607415261097. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX PRO Owners Booklet/RE4TVHP03 Rev 56 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior, Device serial numbers: Prefix "TP" followed by a seven-digit numeric sequence.
- FDA recall number
- Z-1546-2026
- Quantity
- 1,001,534
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX AIR Kit/REA4019-01, Publix TRUE METRIX AIR Kit/REA4020-01, Discount Drug Mart TRUE METRIX AIR Kit/REA4025-01, HEB TRUE METRIX AIR Kit/REA4031-01, SunmarkTRUE METRIX AIR Kit/REA4054-01, Rite AidTRUE METRIX AIR Kit/REA4066-01, Centerwell TRUE METRIX AIR Kit/REA4081-01, HumanaTRUE METRIX AIR Meter Only/REA4081-40, Centerwell TRUE METRIX AIR Starter Kit/REA4081-43, Healthmart TRUE METRIX AIR Kit/REA4089-01, Relion TRUE METRIX AIR Kit/REA4094-01, Relion TRUE METRIX AIR Kit/REA4094-01RX, Kroger TRUE METRIX AIR Kit/REA4203-01, Foster and Thrive TRUE METRIX AIR Kit/REA4211-01, TRUE METRIX AIR Kit/REA4H01-01, TRUE METRIX AIR NFRS Meter Only/REA4H01-40, TRUE METRIX AIR NFRS Starter Kit/REA4H01-43, TRUE METRIX AIR Kit (mg/dL)/REA4i03-01, TRUE METRIX AIR Starter Kit (mg/dL)/REA4i03-02, Walmart (MediMart) TRUE METRIX AIR Meter Only (Mexico - mg/dL)/REA4i04-00, TRUE METRIX AIR Kit (Australia - mmol/L)/REA4i81-11, TRUE METRIX AIR Kit (United Kingdom - mmol/L)/REA4i82-11, TRUE METRIX AIR Starter Kit (United Kingdom - mmol/L)/REA4i82-12
- Lot or code
- Device Item/UDI-DI: REA4001-01/311917173894, REA4002-01/96295129267, REA4007-01/50428354766, REA4011-01/87701427398, REA4019-01/708820685659, REA4020-01/41415072787, REA4025-01/93351022784, REA4031-01/41220009169, REA4054-01/10939742445, REA4066-01/11822590051, REA4081-01/21292015015, REA4081-40/21292007164, REA4081-43/21292007584, REA4089-01/52569139042, REA4094-01/681131240154, REA4094-01RX/681131403214, REA4203-01/41260007354, REA4211-01/10939958617, REA4H01-01/21292007010, REA4H01-40/21292007447, REA4H01-43/21292007911, International: REA4i03-01/N/A, REA4i03-02/N/A, REA4i04-00/21292012731, REA4i81-11/N/A, REA4i82-11/21292012212, REA4i82-12/21292009991. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX AIR Owner's Booklet/ REA4TVH03 Rev 57 and prior, TRUE METRIX AIR Owner's Booklet (English)/ REA4TVH35 Rev 51 and prior, TRUE METRIX AIR Owner's Booklet (Spanish)/ REA4TVH35S Rev 51 and prior, Relion (Walmart) TRUE METRIX AIR Owners Booklet/ REA4RLN03 Rev 52 and prior, Relion (Walmart) TRUE METRIX AIR Owner's Booklet - English/ REA4RLN35 Rev 52 and prior, TRUE METRIX AIR Owner's Booklet International Trividia House Brand/ REA4ITV03 Rev 55 and prior, TRUE METRIX AIR Owners Booklet for Trividia Health UK Limited/ REA4UKT03 Rev 55 and prior, Trividia Health LATAM - TRUE METRIX AIR Owners Booklet/ REA4LAT03 Rev 51 and prior, TRUE METRIX AIR Owners Booklet for Trividia Health Australia/ REA4AUT03 Rev 58 and prior. Device serial numbers: Prefixes "TB" or "TA"…
- FDA recall number
- Z-1547-2026
- Quantity
- 3,678,026
The text above is the beginning of the record; 44 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX GO Kit/RF4001-01BK, CVS TRUE METRIX GO Kit/RF4007-01, Meijer TRUE METRIX GO Kit/RF4019-01, Fred's Phamarcy TRUE METRIX GO Kit/RF4023-01, HEB TRUE METRIX GO Kit/RF4031-01, Hyvee TRUE METRIX GO Kit/RF4048-01, Rite Aid TRUE METRIX GO Kit/RF4066-01, TopCo TRUE METRIX GO Kit/RF4209-01, TRUE METRIX GO Kit/RF4H01-01BK, TRUE METRIX GO NFRS Meter Only/RF4H01-40, TRUE METRIX GO Kit (Jamaica - mmol/L)/RF4i29-11BK, TRUE METRIX GO Kit (Australia - mmol/L)/RF4i81-11BK, TRUE METRIX GO Kit (United Kingdom - mmol/L)/RF4i82-11BK, TRUE METRIX GO Starter Kit (United Kingdom - mmol/L)/RF4i82-12BK
- Lot or code
- Device Item/UDI-DI: RF4001-01BK/311917178691, RF4007-01/50428560402, RF4019-01/708820721098, RF4023-01/21292009335, RF4031-01/41220136131, RF4048-01/75450843576, RF4066-01/11822002783, RF4209-01/36800234567, RF4H01-01BK/21292006112, RF4H01-40/21292009793. International: RF4i29-11BK/N/A, RF4i81-11BK/N/A, RF4i82-11BK/21292012229, RF4i82-12BK/21292009984. Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX GO Owner's Booklet/ RF4TVH03 Rev 57 and prior, TRUE METRIX GO Owner's Booklet (English)/RF4TVH35 Rev 51 and prior, TRUE METRIX GO Owner's Booklet (Spanish)/RF4TVH35S Rev 51 and prior, TRUE METRIX GO Blood Glucose Monitoring System IFU Trividia House Brand/ RF4ITV16 Rev 55 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health UK Limited/ RF4UKT16 Rev 56 and prior, Trividia Health LATAM- TRUE METRIX GO System IFU/ RF4LAT16 Rev 51 and prior, TRUE METRIX GO System Instructions for Use (IFU) for Trividia Health Australia/ RF4AUT16 Rev 57 and prior. Device serial numbers: Prefix "F" followed by an eight-digit numeric sequence.
- FDA recall number
- Z-1548-2026
- Quantity
- 1,407,465
Distribution (FDA): Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.