FDAMedical devicesClass IOngoing

FDA recall 99835: Ventec Life Systems, Inc.: VOCSN+Pro and 11 more

FDA recall 99835 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

37 more codes are in the record below.

  • VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007792, 00855573007006
    FDA recall number
    Z-0122-2027
    Quantity
    4,369 units
  • VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007761, 00855573007204
    FDA recall number
    Z-0123-2027
    Quantity
    325 units
  • VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007785, 00855573007433
    FDA recall number
    Z-0124-2027
    Quantity
    1,534 units
  • VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007211
    FDA recall number
    Z-0125-2027
    Quantity
    606 units
  • VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007747
    FDA recall number
    Z-0126-2027
    Quantity
    1,532 units
  • VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00850018761130
    FDA recall number
    Z-0127-2027
    Quantity
    100 units
  • VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007860
    FDA recall number
    Z-0128-2027
    Quantity
    4,368 units
  • V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.

    Lot or code
    All Serial Numbers: PRT-01185-000 -UDI: 00855573007877; PRT-01185-001 - UDI: 00850018761161; PRT-01185-002 - UDI: 00850018761154
    FDA recall number
    Z-0129-2027
    Quantity
    39,568 units
  • VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007730
    FDA recall number
    Z-0130-2027
    Quantity
    930 units
  • VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00850018761116
    FDA recall number
    Z-0131-2027
    Quantity
    180 units
  • V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007945
    FDA recall number
    Z-0132-2027
    Quantity
    30,120 units
  • V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.

    Lot or code
    All Serial Numbers: UDI: 00855573007822
    FDA recall number
    Z-0133-2027
    Quantity
    25,112 units

Distribution (FDA): Distribution US nationwide including Puerto Rico.

Source

Issued by
U.S. Food and Drug Administration
Firm
Ventec Life Systems, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Ventec Life Systems, Inc., Month: October 2026, Week 2026-W44. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99835 on October 28, 2026.

Other notices of Ventec Life Systems, Inc.