What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
- Lot or code
- REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
- FDA recall number
- Z-2525-2026
- Quantity
- 1
Distribution (FDA): US Nationwide distribution in the state of UT.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Ventec Life Systems, Inc. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.