What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Alphenix INFX-8000C/B5
- Lot or code
- Serial no. B5F2532011/UDI: 4987670100192
- FDA recall number
- Z-0108-2027
- Quantity
- 1 unit
Distribution (FDA): US Nationwide distribution in the state of ID.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Canon Medical System, USA, INC. (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.