Canon Medical System, USA, INC. recalls
7 notices naming Canon Medical System, USA, INC. as the recalling firm, since October 15, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Canon Medical System, USA, INC.: ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C and 2 more
Hazard (FDA): Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
FDAMedical devices
Canon Medical System, USA, INC.: Alphenix INFX-8000C/B5
Hazard (FDA): Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
FDAMedical devices
Canon Medical System, USA, INC.: Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital…
Hazard (FDA): Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
FDAMedical devices
Canon Medical System, USA, INC.: Alphenix INFX-8000C and 3 more
Hazard (FDA): It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result…
FDAMedical devices
Canon Medical System, USA, INC.: VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 and 1 more
Hazard (FDA): There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed…
FDAMedical devices
Canon Medical System, USA, INC.: VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Hazard (FDA): Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
FDAMedical devices
Canon Medical System, USA, INC.: This device is a digital radiography/fluoroscopy system used in a diagnostic and…
Hazard (FDA): It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the…
FDAMedical devices