What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
- Lot or code
- UDI-DI: 00840682102292; Serial Numbers: APWC01722 APWF00169 APWF00170 APWF00173 APWF00174 APWF00111 APWF00368 APWF00402 APWF00465 APWF00467 APWF00468 APWF00469 APWF00470 APWF00471 APWF00472 APWF00473 APWF00474 APWF00475 APWF00476 APWF00477 APWT00394 APWC01107 APWF00436 APWE01910 APWE01911 APWV00635 APWW01464 APWW01625 APWF00380 APWF00381 APWF00382 APWY00358 APWF00359 APWF00313 APWE01925 APWV02342 APWE01915 APWF00458 APWF00433 APWD02646 APWC01330 APWC01339 APWC01340 APWF00279 APWF00280 APWF00281 APWF00282 APWF00283 APWF00284 APWF00285 APWF00286 APWF00287 APWF00288 APWF00289 APWF00290 APWF00262 APWF00312 APWX00800 APWX00804 APWF00138 APWF00466 APWE01974 APWF00396 APWF00297 APWF00298 APWF00299 APWF00300 APWF00296 APWT00798 APWA02035 APWF00239 APWF00241 APWF00242 APWF00243 APWF00244 APWF00245 APWF00246 APWF00247 APWF00248 APWF00249 APWF00250 APWF00251 APWF00252 APWF00253 APWF00254 APWF00255 APWF00256 APWF00405 APWF00406 APWF00459 APWF00145 APWF00146 APWF00147 APWF00149 APWF00150 APWF00346 APWE02051 APWE02057 APWE02058 APWE02059 APWE01912 APWF00308 APWF00309 APWE01990 APWW00228 APWU01342 APWE01989 APWT00240 APWZ00855 APWF00395 APWF00339 APWX02985 APWX02993 APWF00190 APWE01968 APWC01311 APWF00456 APWY01027 APWY01033 APWY01034 APWE01960 APWF00197 APWE01972 APWE01973 APWC01482 APWF00304 APWE01933 APWE01935 APWF00460 APWZ01205 APWF00237 APWE01998 APWE01999 APWE01971 APWF00179 APWF00180 APWF00181 APWF00182 APWF00183 APWD02593 APWD02605 APWC03541 APWF00294 APWF00311 APWV02707 APWV02725…
- FDA recall number
- Z-3189-2026
- Quantity
- 475 units
The text above is the beginning of the record; 3,290 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000
- Lot or code
- UDI-DI: 00195278588128; Serial Numbers: AQDF40422 AQDE40769 AQDE40771 AQDF40287 AQDF40288 AQDF40290 AQDF40291 AQDF40293 AQDF40294 AQDF40295 AQDF40311 AQDF40313 AQDF40314 AQDE40754 AQDE40755 AQDE40756 AQDE40757 AQDE40759 AQDE40762 AQDE40763 AQDF40296 AQDF40300 AQDF40301 AQDF40302 AQDF40303 AQDD40441 AQDF40399 AQDF40400 AQDF40401 AQDF40423 AQDF40424 AQDF40425 AQDE40743 AQDE40744 AQDE40748 AQDE40749 AQDC40064 AQDE40427 AQDC40059 AQDE40736 AQDF40353 AQDF40354 AQDF40355 AQDF40356 AQDF40357 AQDF40358 AQDF40359 AQDF40360 AQDF40361 AQDF40362 AQDF40363 AQDF40364 AQDF40365 AQDF40366 AQDF40367 AQDF40368 AQDF40369 AQDE40780 AQDE40781 AQDE40782 AQDE40783 AQDE40787 AQDE40788 AQDF40325 AQDF40371 AQDF40372 AQDF40373 AQDE40753 AQDF40312 AQDF40382 AQDF40383 AQDF40384 AQDF40385 AQDF40386 AQDF40387 AQDF40308 AQDF40309 AQDF40310 AQDE40765 AQDE40790 AQDE40792 AQDF40398 AQDF40307 AQDE40766 AQDE40767 AQDE40778 AQDE40770 AQDE40774 AQDE40775 AQDE40776 AQDE40777 AQDE40058 AQDD40417 AQDE40799 AQDE40764 AQDE40800 AQDF40283 AQDF40316 AQDF40317 AQDF40318 AQDF40319 AQDF40320 AQDF40321 AQDF40322 AQDF40323 AQDF40324 AQDF40327 AQDF40414 AQDE40739 AQDE40741 AQDE40742 AQDE40779 AQDF40388 AQDF40389 AQDF40390 AQDF40391 AQDE40737 AQDF40350 AQDF40351 AQDF40352 AQDF40374 AQDF40375 AQDF40376 AQDF40377 AQDF40378 AQDF40379 AQDF40380 AQDF40304 AQDF40381 AQDF40415 AQDF40416 AQDF40417 AQDF40418 AQDF40419 AQDF40420 AQDF40421 AQDE40731 AQDE40732 AQDE40735 AQDF40326
- FDA recall number
- Z-3190-2026
- Quantity
- 140 units
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
- Lot or code
- Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
- FDA recall number
- Z-3191-2026
- Quantity
- 7 units
Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665
- Lot or code
- Serial Numbers: 801042711 801031045 801031926 801045423 801011743 801090568 801128677 801011276-1 801011276-10 801011276-11 801011276-12 801011276-13 801011276-14 801011276-15 801011276-2 801011276-3 801011276-4 801011276-5 801011276-6 801011276-7 801011276-8 801011276-9 801078342-1 801078342-10 801078342-11 801078342-12 801078342-13 801078342-14 801078342-15 801078342-16 801078342-17 801078342-18 801078342-19 801078342-2 801078342-20 801078342-21 801078342-22 801078342-23 801078342-24 801078342-25 801078342-26 801078342-27 801078342-28 801078342-29 801078342-3 801078342-30 801078342-4 801078342-5 801078342-6 801078342-7 801078342-8 801078342-9 801102924 801108227 801113705 801114965 801084897 801003006 801004394 801071735 801016238 801050584 801051585 801064585 801013238 801013095 801043259 801088072 801014199 801006762 801052860 801023283 801110573 801064058 801080036 801126195 801064184-1 801064184-2 801059320-1 801059320-2 801091614 801028252 801051344 801030002 800988180 801111226 801028748 801054482 801090144 801083490 801105951 107945491 801101919 801121340 801038810 801053994 801105562 801119480 801049728 801078062-1 801078062-2 801089690-1 801089690-2 801115703-1 801115703-2 801016353 801019099 801041178 801052685 801081089 801126428 800998007-1 800998007-2 801078509-1 801078509-2 801118676-1 801118676-2 801096089 801101334-1 801101334-2 801100493 801039330 801041393 801055432 801119046 801115340 801040296 801111527 801009694 801030435 801061340 801068616 801044691…
- FDA recall number
- Z-3192-2026
- Quantity
- 410 units
The text above is the beginning of the record; 2,909 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide - US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Datex-Ohmeda, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.