Skip to content
FDAMedical devicesClass IOngoing

FDA recall 99036: Datex-Ohmeda Inc: Giraffe Stand-alone Infant Resuscitation System and 2 more

FDA recall 99036 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

  • Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance

    Lot or code
    UDI-DI of service blender component: 00813841020010 Affected devices were installed with the following serial numbers of affected service blender components: 002I08B 002X08B 008S10B 008Y09B 013S09B 016A09B 016R09B 018P10B 019J09B 028C08B 02X08B 033N09B 037E08B 038D10B 038N08B 042F10B 042S10B 042T10B 045808B 045M08B 047I08B 047J08B 054C10B 054I10B 054L10B 054N09B 054O09B 054U10B 055D09B 055V09B 057908B 061K09B 062C10B 068W10B 077M09B 078509B 078M09B 078S09B 084I10B 084J10B 092G08B 099F08B 1 B1405629 100E08B 100R08B 104A09B 12105790 133D08B 136708B 136W08B 141J09B 142009B 142W09B 144J08B 144N08B 149Q09B 163608B 169D08B 169G08B 169H08B 172308B 175P08B 185I08B 185Q08B 186208B 189808B 189A09B 189C08B 189F09B 193L08B 193N08B 195D09B 195E09B 195J09B 195K09B 198H08B 210K09B 210U09B 218N09B 219C09B 219K09B 220E08B 226F08B 244L09B 260X09B 266M08B 267808B 280F09B 281H09B 285J07B 295H09B 295J09B 295P08B 310I09B 338707B 338E07B 338K07B 352Y09B 355P09B 7979700385 B1003093 B1003112 B1003124 B1003140 B1003143 B1003147 B1003153 B1003343 B1003452 B1003550 B1003867 B1003944 B1004844 B1004861 B1004924 B1005029 B1005036 B1005044 B1005422 B1005534 B1005560 B1005571 B1005839 B1006102 B1006113 B1006304 B1006858 B1006935 B1006971 B1007078 B1007220 B1007262 B1007355 B1007360 B1007376 B1007383 B1007528 B1007605 B1101215 B1101224 B1101241 B1101445 B1101446 B1101547 B1101862 B1101869 B1101998 B1102011 B1102015 B1102153 B1102384 B1102405 B1102571 B1103145 B1103159 B1103251 B1103256 B1103422 B1103434…
    FDA recall number
    Z-2357-2026
    Quantity
    550 (367 US; 183 OUS)

    The text above is the beginning of the record; 23,841 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

    Lot or code
    UDI-DI of service blender component: 00813841020010 Affected devices were installed with the following serial numbers of affected service blender components: 002I08B 002X08B 008S10B 008Y09B 013S09B 016A09B 016R09B 018P10B 019J09B 028C08B 02X08B 033N09B 037E08B 038D10B 038N08B 042F10B 042S10B 042T10B 045808B 045M08B 047I08B 047J08B 054C10B 054I10B 054L10B 054N09B 054O09B 054U10B 055D09B 055V09B 057908B 061K09B 062C10B 068W10B 077M09B 078509B 078M09B 078S09B 084I10B 084J10B 092G08B 099F08B 1 B1405629 100E08B 100R08B 104A09B 12105790 133D08B 136708B 136W08B 141J09B 142009B 142W09B 144J08B 144N08B 149Q09B 163608B 169D08B 169G08B 169H08B 172308B 175P08B 185I08B 185Q08B 186208B 189808B 189A09B 189C08B 189F09B 193L08B 193N08B 195D09B 195E09B 195J09B 195K09B 198H08B 210K09B 210U09B 218N09B 219C09B 219K09B 220E08B 226F08B 244L09B 260X09B 266M08B 267808B 280F09B 281H09B 285J07B 295H09B 295J09B 295P08B 310I09B 338707B 338E07B 338K07B 352Y09B 355P09B 7979700385 B1003093 B1003112 B1003124 B1003140 B1003143 B1003147 B1003153 B1003343 B1003452 B1003550 B1003867 B1003944 B1004844 B1004861 B1004924 B1005029 B1005036 B1005044 B1005422 B1005534 B1005560 B1005571 B1005839 B1006102 B1006113 B1006304 B1006858 B1006935 B1006971 B1007078 B1007220 B1007262 B1007355 B1007360 B1007376 B1007383 B1007528 B1007605 B1101215 B1101224 B1101241 B1101445 B1101446 B1101547 B1101862 B1101869 B1101998 B1102011 B1102015 B1102153 B1102384 B1102405 B1102571 B1103145 B1103159 B1103251 B1103256 B1103422 B1103434…
    FDA recall number
    Z-2358-2026
    Quantity
    550 (367 US; 183 OUS)

    The text above is the beginning of the record; 23,841 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.

    Lot or code
    UDI-DI of service blender component: 00813841020010 Affected devices were installed with the following serial numbers of affected service blender components: 002I08B 002X08B 008S10B 008Y09B 013S09B 016A09B 016R09B 018P10B 019J09B 028C08B 02X08B 033N09B 037E08B 038D10B 038N08B 042F10B 042S10B 042T10B 045808B 045M08B 047I08B 047J08B 054C10B 054I10B 054L10B 054N09B 054O09B 054U10B 055D09B 055V09B 057908B 061K09B 062C10B 068W10B 077M09B 078509B 078M09B 078S09B 084I10B 084J10B 092G08B 099F08B 1 B1405629 100E08B 100R08B 104A09B 12105790 133D08B 136708B 136W08B 141J09B 142009B 142W09B 144J08B 144N08B 149Q09B 163608B 169D08B 169G08B 169H08B 172308B 175P08B 185I08B 185Q08B 186208B 189808B 189A09B 189C08B 189F09B 193L08B 193N08B 195D09B 195E09B 195J09B 195K09B 198H08B 210K09B 210U09B 218N09B 219C09B 219K09B 220E08B 226F08B 244L09B 260X09B 266M08B 267808B 280F09B 281H09B 285J07B 295H09B 295J09B 295P08B 310I09B 338707B 338E07B 338K07B 352Y09B 355P09B 7979700385 B1003093 B1003112 B1003124 B1003140 B1003143 B1003147 B1003153 B1003343 B1003452 B1003550 B1003867 B1003944 B1004844 B1004861 B1004924 B1005029 B1005036 B1005044 B1005422 B1005534 B1005560 B1005571 B1005839 B1006102 B1006113 B1006304 B1006858 B1006935 B1006971 B1007078 B1007220 B1007262 B1007355 B1007360 B1007376 B1007383 B1007528 B1007605 B1101215 B1101224 B1101241 B1101445 B1101446 B1101547 B1101862 B1101869 B1101998 B1102011 B1102015 B1102153 B1102384 B1102405 B1102571 B1103145 B1103159 B1103251 B1103256 B1103422 B1103434…
    FDA recall number
    Z-2359-2026
    Quantity
    550 (367 US; 183 OUS)

    The text above is the beginning of the record; 23,841 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Pending

Source

Issued by
U.S. Food and Drug Administration
Firm
Datex-Ohmeda Inc
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Datex-Ohmeda Inc, Month: June 2026, Week 2026-W26. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99036 on June 24, 2026.

Other notices of Datex-Ohmeda Inc