Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99617: Percussionaire Corporation: Intrapulmonary Percussive Ventilation and 6 more

FDA recall 99617 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

14 more codes are in the record below.

  • Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual

    Lot or code
    Product Number/UDI-DI: S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev D
    FDA recall number
    Z-3319-2026
    Quantity
    61,858
  • In-Line Valve, Product Number: P5-TEE

    Lot or code
    UDI: 00849436000723
    FDA recall number
    Z-3320-2026
    Quantity
    98,434
  • Intrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual

    Lot or code
    Product Number/UDI-DI: S00002-C/00849436000020, 00849436000563, 00849436000570, 00849436000143, 00849436000648; P5/00849436000501; Instruction Manual: P20045 Rev B
    FDA recall number
    Z-3321-2026
    Quantity
    229,105
  • Intrapulmonary Percussive Ventilation, IPV-2C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual

    Lot or code
    Product Number/UDI-DI: S00002-C/00849436000020, 00849436000563, 00849436000570, 00849436000143, 00849436000648; P5/00849436000501; Instruction Manual: P20046 Rev A
    FDA recall number
    Z-3322-2026
    Quantity
    229,430
  • Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual

    Lot or code
    Product Number/UDI-DI: S00010/00849436000389, P5/00849436000501, Instruction Manual: P20047 Rev E
    FDA recall number
    Z-3323-2026
    Quantity
    228,277
  • Travel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual

    Lot or code
    Product Number/UDI-DI: S00055/00849436000242, P5/00849436000501, Instruction Manual: P20106 Rev B
    FDA recall number
    Z-3324-2026
    Quantity
    228,596
  • Intrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC

    Lot or code
    Product/Catalog Number/UDI-DI: S00009/00849436000358, P5-HC/00849436000525
    FDA recall number
    Z-3325-2026
    Quantity
    419

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.

Source

Issued by
U.S. Food and Drug Administration
Firm
Percussionaire Corporation
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Percussionaire Corporation, Month: October 2026, Week 2026-W41. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99617 on October 7, 2026.