Percussionaire Corporation recalls
3 notices naming Percussionaire Corporation as the recalling firm, since January 29, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Percussionaire Corporation: Intrapulmonary Percussive Ventilation and 6 more
Hazard (FDA): Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents…
FDAMedical devices
Percussionaire Corporation: Phasitron 5
Hazard (FDA): IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of…
FDAMedical devices
Percussionaire Corporation: VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK
Hazard (FDA): Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
FDAMedical devices