What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
2 more codes are in the record below.
StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.
- Lot or code
- Model No: 63685. UDI-DI: 00385480636858. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
- FDA recall number
- Z-3128-2026
- Quantity
- 20,616 units
StatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.
- Lot or code
- Model No: 66739. UDI-DI: 0038548063685. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
- FDA recall number
- Z-3129-2026
- Quantity
- 10,941 units
StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.
- Lot or code
- Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
- FDA recall number
- Z-3130-2026
- Quantity
- 1,858 units
StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.
- Lot or code
- Model No: 63910. UDI-DI: 00385480639101. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
- FDA recall number
- Z-3131-2026
- Quantity
- 8,939 units
StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.
- Lot or code
- Model No: 66429. UDI-DI: :00385480664295. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
- FDA recall number
- Z-3132-2026
- Quantity
- 2,044 units
Distribution (FDA): Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Nova Biomedical Corporation
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.