Nova Biomedical Corporation recalls
2 notices naming Nova Biomedical Corporation as the recalling firm, since January 1, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Nova Biomedical Corporation: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system. and 4 more
Hazard (FDA): A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID)…
FDAMedical devices
Nova Biomedical Corporation: StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use… and 2 more
Hazard (FDA): A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using…