What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900
- Lot or code
- Flow-i C20 6677200 / UDI-DI 07325710001349 Flow-i C30 6677300 / UDI-DI 07325710001363 Flow-i C40 6677400 / UDI-DI 07325710001318 Flow-i C20 6888520 / UDI-DI 07325710010617 Flow-i C30 6888530 / UDI-DI 07325710010624 Flow-i C40 6888540 / UDI-DI 07325710010631 Flow-c 6887700 / UDI-DI 07325710009765 Flow-e 6887900 / UDI-DI 07325710010457 Software versions 4.8, 4.10, and 4.11.
- FDA recall number
- Z-2860-2026
- Quantity
- 408 US; 11,342 OUS
Distribution (FDA): Worldwide - US Nationwide distribution to the states of AL, AR, CA, CO, DC, DE, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, SC, TX, VA, WA, WV.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Maquet Critical Care AB