What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Servo-u Ventilator System. Model Number: 6688600.
- Lot or code
- Model Number: 6688600. UDI-DI: 07325710001042. All software versions.
- FDA recall number
- Z-2013-2025
- Quantity
- 874 units
Servo-n Ventilator System. Model Number: 6694800.
- Lot or code
- Model Number: 6694800. UDI-DI: 07325710001110. All software versions.
- FDA recall number
- Z-2014-2025
- Quantity
- 7,688 units
Servo-u MR Ventilator System. Model Number: 6888800.
- Lot or code
- Model Number: 6888800. UDI-DI: 07325710010518. All software versions.
- FDA recall number
- Z-2015-2025
- Quantity
- 129 units
Distribution (FDA): US Nationwide distribution and Puerto Rico. No international distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Maquet Critical Care AB