What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump
- Lot or code
- Software application (app) version 1.x and 2.x, loaded on the following units: 1. Model Number CT900A, GTIN 00643169890923, Serial Numbers: NPL0004445; 2. Model Number CT900B, GTIN 00643169890947, Serial Numbers: NPL1000042, NPL1000045, NPL1000046, NPL1000048, NPL1000050, NPL1000051, NPL1000053, NPL1000054, NPL1000059, NPL1000060, NPL1000062, NPL1000065, NPL1000069, NPL1000070, NPL1000071, NPL1000072, NPL1000073, NPL1000074, NPL1000075, NPL1000077, NPL1000081, NPL1000082, NPL1000083, NPL1000085, NPL1000086, NPL1000087, NPL1000090, NPL1000091, NPL1000093, NPL1000094, NPL1000098, NPL1000117, NPL1000133, NPL1000135, NPL1000137, NPL1000138, NPL1000151, NPL1000159, NPL1000160, NPL1000164, NPL1000166, NPL1000167, NPL1000169, NPL1000173, NPL1000174, NPL1000179, NPL1000180, NPL1000183, NPL1000185, NPL1000186, NPL1000187, NPL1000188, NPL1000191, NPL1000192, NPL1000193, NPL1000194, NPL1000195, NPL1000196, NPL1000230, NPL1000235, NPL1000237, NPL1000242, NPL1000244, NPL1000245, NPL1000246, NPL1000248, NPL1000249, NPL1000252, NPL1000254, NPL1000257, NPL1000258, NPL1000281, NPL1000292, NPL1000297, NPL1000301, NPL1000303, NPL1000307, NPL1000321, NPL1000322, NPL1000323, NPL1000325, NPL1000326, NPL1000332, NPL1000334, NPL1000343, NPL1000361, NPL1000362, NPL1000363, NPL1000364, NPL1000365, NPL1000366, NPL1000367, NPL1000369, NPL1000370, NPL1000381, NPL1000483, NPL1000486, NPL1000487, NPL1000501, NPL1000502, NPL1000505, NPL1000506, NPL1000507, NPL1000509, NPL1000510, NPL1000512, NPL1000515,…
- FDA recall number
- Z-2201-2026
- Quantity
- 10165 units
The text above is the beginning of the record; 121,146 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medtronic Neuromodulation
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.