Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 98028: Medtronic Neuromodulation: Restore Clinician Programmer Application

FDA recall 98028 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG

    Lot or code
    CP App: UDI-DI 00763000273668 all software versions prior to v.1.0.4489 are affected The implanted devices are not affected.
    FDA recall number
    Z-0996-2026
    Quantity
    7667 devices

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Reunion, Russia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medtronic Neuromodulation
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medtronic Neuromodulation, Month: December 2025, Week 2026-W01. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 98028 on December 31, 2025.

Other notices of Medtronic Neuromodulation