What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Brand Name: Leksell Vantage Arc System REF: 1053958
- Lot or code
- Lot Code: Arc System: UDI:07340048304856 Parts of Arc System: Left Z-Slide: Serial Numbers :UDI: 07340048304818 00001 00625 Right Z-Slide: UDI: 07340048304801/ Serial Numbers: 00001 00675
- FDA recall number
- Z-2006-2026
- Quantity
- 768 (364Left/364Right)
Distribution (FDA): US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea, Republic of Kuwait Lithuania Mexico Morocco Netherlands New Zealand Norway Pakistan Panama Poland Portugal Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Turkey United Kingdom
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Elekta, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.