Elekta, Inc. recalls
9 notices naming Elekta, Inc. as the recalling firm, since February 19, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Elekta, Inc.: Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote… and 2 more
Hazard (FDA): Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.
FDAMedical devices
Elekta, Inc.: MOSAIQ Oncology Information System
Hazard (FDA): Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
FDAMedical devices
Elekta, Inc.: Brand Name: Leksell Vantage Arc System REF: 1053958
Hazard (FDA): Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
FDAMedical devices
Elekta, Inc.: Brand Name: Leksell Gamma Knife Product Description: Leksell Gamma Knife¿ Perfexion…
Hazard (FDA): Due to adapter not being properly aligned or locked in place.
FDAMedical devices
Elekta, Inc.: Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system… and 1 more
Hazard (FDA): Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates…
FDAMedical devices
Elekta, Inc.: Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy
Hazard (FDA): Due to a manufacturing issue and electrical grounding of systems.
FDAMedical devices
Elekta, Inc.: MOSAIQ Oncology Information System with Particle Therapy License
Hazard (FDA): Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial…
FDAMedical devices
Elekta, Inc.: Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic…
Hazard (FDA): Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
FDAMedical devices
Elekta, Inc.: MOSAIQ Oncology Information System
Hazard (FDA): Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.